MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/jcm9041086

Медицинская библиотека · Journal of clinical medicine · 2020

A New System for Assessing Visual Disability Using a Digital Visor.

Sangiuolo Raffaele, Amore Filippo, Bacci Mauro, Brusini Paolo, Cruciani Filippo, Gualtieri Giacomo, Lancia Massimo, Sangiuolo Giulia, Sangiuolo Mario

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Considering the lack of universally accepted visual requirements for driving and for defining various grades of visual disability, the aim of this study is to propose a new method that provides a numerical score resulting from a combined assessment of the visual field and visual acuity loss obtained using a digital technology visor.

METHODS

This study presents a new system for calculating the percentage of visual disability by combining binocular visual acuity and binocular visual field assessments. A new Global Vision Evaluation System digital technology visor uses standardized, reproducible criteria to produce well-defined, numerically expressed test results. Through a specific algorithm, the device produces a numerical value expressing the percentage of visual disability.

RESULTS

Eighty-six subjects with various types of visual impairment underwent visual acuity and visual field test examinations carried out employing both traditional methods and the new digital visor. The two methods provided homogeneously similar results regarding the positioning of the subjects on the visual disability scale.

CONCLUSIONS

The new digital visor seems to be a valid method to ensure that visual disability assessments are more homogeneous and reliable, and that, consequently, the resources available for this purpose are more fairly distributed.

Для цитирования: Sangiuolo Raffaele, Amore Filippo, Bacci Mauro, Brusini Paolo, Cruciani Filippo, Gualtieri Giacomo, Lancia Massimo, Sangiuolo Giulia, Sangiuolo Mario. A New System for Assessing Visual Disability Using a Digital Visor.. Journal of clinical medicine. 2020. DOI: 10.3390/jcm9041086.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.