A Haptic Guided Robotic System for Endoscope Positioning and Holding.

Cabuk Burak, Ceylan Savas, Anik Ihsan, Tugasaygi Mehtap, Kizir Selcuk

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIM: To determine the feasibility, advantages, and disadvantages of using a robot for holding and maneuvering the endoscope in transnasal transsphenoidal surgery.

MATERIAL AND

METHODS

The system used in this study was a Stewart Platform based robotic system that was developed by Kocaeli University Department of Mechatronics Engineering for positioning and holding of endoscope. After the first use on an artificial head model, the system was used on six fresh postmortem bodies that were provided by the Morgue Specialization Department of the Forensic Medicine Institute (Istanbul, Turkey).

RESULTS

The setup required for robotic system was easy, the time for registration procedure and setup of the robot takes 15 minutes. The resistance was felt on haptic arm in case of contact or friction with adjacent tissues. The adaptation process was shorter with the mouse to manipulate the endoscope. The endoscopic transsphenoidal approach was achieved with the robotic system. The endoscope was guided to the sphenoid ostium with the help of the robotic arm.

CONCLUSION

This robotic system can be used in endoscopic transsphenoidal surgery as an endoscope positioner and holder. The robot is able to change the position easily with the help of an assistant and prevents tremor, and provides a better field of vision for work.

Для цитирования: Cabuk Burak, Ceylan Savas, Anik Ihsan, Tugasaygi Mehtap, Kizir Selcuk. A Haptic Guided Robotic System for Endoscope Positioning and Holding.. Turkish neurosurgery. 2015. DOI: 10.5137/1019-5149.jtn.13290-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.