OBJECTIVE

The goal of this study was to examine the theoretical feasibility of a new manipulator system for endoscope guidance in functional endoscopic sinus surgery.

STUDY

DESIGN

The accuracy of endoscope positioning and time of endoscope movement with an endoscope manipulator system were determined with an artificial sinus model.

SETTING

A laboratory trial was performed. The time for 60 repetitions of manual compared to manipulator-assisted endoscope movements directed at 3 different target positions was evaluated. In addition, the alignment of the position vector for each endoscope movement was examined.

SUBJECTS AND

METHODS

A zero-degree Hopkins II telescope with a camera was used to head for the target positions. First, the endoscope movements were done manually, and afterward the endoscope manipulator system was used for endoscope guidance. The alignment of the position vector of the endoscope was measured with a portable measuring arm.

RESULTS

There was no statistical difference between the time for manual and manipulator-assisted endoscope movements for all target positions. The alignment of the position vector of the endoscope was statistically different at 2 target positions: anterior ethmoid left side and ostium of maxillary sinus left side. There was no statistical difference at all other positions.

CONCLUSION

The endoscope manipulator system has the potential to be integrated into the operating workflow without extending the time for endoscope guidance. The surgeon will be able to use both hands for the manipulation of the instruments. Less frequent endoscope movements and instrument changes may be expected after technical modification.

Для цитирования: Fischer Milos, Gröbner Christina, Dietz Andreas, Krinninger Maximillian, Lüth Tim C, Strauss Gero. A technique with manipulator-assisted endoscope guidance for functional endoscopic sinus surgery: proof of concept.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2011. DOI: 10.1177/0194599811412726.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.