MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1148/radiology.214.1.r00ja42107

Медицинская библиотека · Radiology · 2000

MR hysterosalpingography: protocol development and refinement for simulating normal and abnormal fallopian tube patency--feasibility study with a phantom.

Frye R E, Ascher S M, Thomasson D

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To develop and refine a pulse sequence and protocol for testing the feasibility of magnetic resonance (MR) hysterosalpingography in a phantom model.

MATERIALS AND

METHODS

A phantom simulating the uterus, fallopian tubes, and surrounding pelvic cavity was constructed. T2-weighted acquisition strategies-breath-hold fast spin-echo, rapid acquisition with relaxation enhancement (RARE), and haff-Fourier RARE-were refined to acquire sequential 70-mm coronal imaging volumes. Contrast agent was injected into the introducing catheter entering the os of the simulated uterus. Interacquisition interval, type of contrast agent (eg, sterile saline solution or water), and quantity of contrast agent (eg, 1-5 mL per acquisition) were varied. Digital image subtraction was used to enhance image quality. Images were qualitatively analyzed and rated good, fair, or poor for temporal resolution, spatial resolution, fallopian tube conspicuity, and free spill conspicuity. Once the technique was refine, the phantom was reconfigured to simulate unilateral and bilateral hydrosalpinx.

RESULTS

The RARE sequence with an 8-second interacquisition interval and a 5-mL interacquisition of sterile water produced good images of the simulated fallopian tubes and free spill. Depiction of unilateral and bilateral hydrosalpinx was also reliably demonstrated.

CONCLUSION

This study with a phantom model demonstrates the feasibility of MR hysterosalpingography to depict normal and diseased fallopian tubes.

Для цитирования: Frye R E, Ascher S M, Thomasson D. MR hysterosalpingography: protocol development and refinement for simulating normal and abnormal fallopian tube patency--feasibility study with a phantom.. Radiology. 2000. DOI: 10.1148/radiology.214.1.r00ja42107.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.