The force required to remove the frameless 0-suture IUD anchoring system: comparison between pre- and postmenopausal women.
Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wildemeersch Dirk
Аннотация
OBJECTIVE: The objective of this short communication is to measure and compare the force needed to remove the implanted "0-suture" GyneFix intrauterine contraceptive device (IUD) and the "0-suture" FibroPlant trade mark -LNG intrauterine system (IUS) from the uterus of pre- and postmenopausal women.
STUDY DESIGN: A nonrandomized comparative study in 119 pre- and postmenopausal women. A dynamometer (Pesola) was used to measure the removal force in newtons.
RESULTS: The results of this study show a mean removal force of 8.5 and 9.5 newtons, respectively (range, 3-11 and 4.5-11), in pre- and postmenopausal women, which is significantly different (p = 0.003).
CONCLUSIONS: The force needed to remove the IUD/IUS anchored in the myometrium of the uterine fundus of pre- and postmenopausal women is higher than the removal force found in previous studies in which the IUD consisted of a slightly thinner anchoring thread (00 instead of 0 suture). The statistically significantly different removal force between the two groups has no clinical implications. The difference may reflect the increased compactness of the uterine tissue in the postmenopausal uterus.
STUDY DESIGN: A nonrandomized comparative study in 119 pre- and postmenopausal women. A dynamometer (Pesola) was used to measure the removal force in newtons.
RESULTS: The results of this study show a mean removal force of 8.5 and 9.5 newtons, respectively (range, 3-11 and 4.5-11), in pre- and postmenopausal women, which is significantly different (p = 0.003).
CONCLUSIONS: The force needed to remove the IUD/IUS anchored in the myometrium of the uterine fundus of pre- and postmenopausal women is higher than the removal force found in previous studies in which the IUD consisted of a slightly thinner anchoring thread (00 instead of 0 suture). The statistically significantly different removal force between the two groups has no clinical implications. The difference may reflect the increased compactness of the uterine tissue in the postmenopausal uterus.