MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11604-015-0471-9

Медицинская библиотека · Japanese journal of radiology · 2015

Added value of a single-energy projection-based metal-artifact reduction algorithm for the computed tomography evaluation of oral cavity cancers.

Hirata Kenichiro, Utsunomiya Daisuke, Oda Seitaro, Kidoh Masafumi, Funama Yoshinori, Yuki Hideaki, Yoshida Morikatsu, Yamashita Yasuyuki

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The purpose of our study was to investigate the added value of the single-energy projection-based metallic artifact reduction (SEMAR) technique on the visualization and diagnostic confidence of oral cavity cancers.

MATERIALS AND

METHODS

Contrast-enhanced CT was performed on 40 patients with dental metalwork. Of these, 18 patients had pathologically confirmed oral cavity cancer, and the other 22 patients had no lesions. CT attenuation and noise in the oral cavity were compared between the conventional and SEMAR images. Two radiologists visually graded the diagnostic confidence on a 5-point scale from 1 (definitely absent) to 5 (definitely present). The value of the SEMAR was assessed in a receiver-operating characteristic curve analysis.

RESULTS

The artifact of the SEMAR images was significantly lower than that of the conventional images (38.4 ± 18.0 HU vs. 187.7 ± 162.7 HU; P < 0.01). Reviewers 1 and 2 detected 10 and 7 more lesions with the SEMAR, respectively. Areas under the curve for the conventional and SEMAR reconstruction were 0.761 and 0.942, respectively, for reviewer 1 and 0.701 and 0.864, respectively, for reviewer 2.

CONCLUSION

The SEMAR significantly improved the oral cavity cancer visualization and diagnostic confidence.

Для цитирования: Hirata Kenichiro, Utsunomiya Daisuke, Oda Seitaro, Kidoh Masafumi, Funama Yoshinori, Yuki Hideaki, Yoshida Morikatsu, Yamashita Yasuyuki. Added value of a single-energy projection-based metal-artifact reduction algorithm for the computed tomography evaluation of oral cavity cancers.. Japanese journal of radiology. 2015. DOI: 10.1007/s11604-015-0471-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.