DOI 10.1007/s00268-023-07082-6
Efficacy and Safety of Gastrointestinal Tumour Site Marking with da Vinci-Compatible Near-Infrared Fluorescent Clips: A Case Series.
Аннотация
BACKGROUND
The conventional near-infrared fluorescent clip (NIRFC) ZEOCLIP FS® has been used successfully in marking tumour sites during laparoscopic surgeries. However, this clip is difficult to observe with the Firefly imaging system equipped with the da Vinci® surgical system. We have been involved in the modification of ZEOCLIP FS® and development of da Vinci-compatible NIRFC. This is the first prospective single-centre case series study verifying the usefulness and safety of the da Vinci-compatible NIRFC.
METHODS
Twenty-eight consecutive patients undergoing da Vinci®-assisted surgery for gastrointestinal cancer (16 gastric, 4 oesophageal, and 8 rectal cases) between May 2021 and May 2022 were enrolled.
RESULTS
Tumour location was identified by the da Vinci-compatible NIRFCs in 21 of 28 (75%) patients, which involved 12 gastric (75%), 4 oesophageal (100%), and 5 rectal (62%) cancer cases. No adverse events were observed.
CONCLUSION
Tumour site marking with da Vinci-compatible NIRFC was feasible in 28 patients enrolled in this study. Further studies are warranted to substantiate the safety and improve the recognition rate.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.