Carbon dioxide diffusion coefficient in noninvasive high-frequency oscillatory ventilation.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Schäfer Christoph, Schumann Stefan, Fuchs Hans, Klotz Daniel
Аннотация
OBJECTIVES: The carbon dioxide (CO2 ) diffusion coefficient (DCO 2 ) reflects CO 2 removal during high-frequency oscillatory ventilation (HFOV). We hypothesized that despite leak flow during noninvasive HFOV (nHFOV) DCO 2 continues to indicate ventilation efficacy.
METHODS: A neonatal airway model including CO2 production and an adjustable oropharyngeal leak was connected to a ventilator via bi-nasal prongs. Pressures and gas flows were measured at prongs, trachea, and leak. Oscillatory tidal volumes below (V T trachea ) and above the leak (V T prong ) were calculated from tracheal and leak flows. DCO 2 was calculated using V T trachea (DCO 2 trachea ) and V T prong (DCO 2 prong ) and compared with CO 2 partial pressure (pCO 2 ). Effects of leak flow (0, 5, or 10 L/min) on DCO 2 were assessed at fixed pressure amplitudes or predefined oscillatory volumes under steady-state pCO 2 conditions in the modeled lung.
RESULTS: DCO2 trachea correlated strongly with pCO 2 , independent of the leak flow level (P < 0.0001). DCO 2 prong correlated with pCO 2 without and with moderate leak (P < 0.0001) but not with maximum leak (P = 0.1432). V T trachea correlated with the quotient of tracheal pressure amplitude and frequency irrespective of the leak (P < 0.0001). Based on the pressure amplitude at prong level (A prong ) V T trachea continued to follow a linear model of which the slopes decreased with increasing leak flow. V T prong correlated with the quotient of A prong and frequency, irrespective of the leak (P < 0.0001).
CONCLUSIONS: DCO2 obtained at the airway opening at prong level reflects ventilation efficacy during nHFOV even in the presence of moderate oropharyngeal leak.
METHODS: A neonatal airway model including CO2 production and an adjustable oropharyngeal leak was connected to a ventilator via bi-nasal prongs. Pressures and gas flows were measured at prongs, trachea, and leak. Oscillatory tidal volumes below (V T trachea ) and above the leak (V T prong ) were calculated from tracheal and leak flows. DCO 2 was calculated using V T trachea (DCO 2 trachea ) and V T prong (DCO 2 prong ) and compared with CO 2 partial pressure (pCO 2 ). Effects of leak flow (0, 5, or 10 L/min) on DCO 2 were assessed at fixed pressure amplitudes or predefined oscillatory volumes under steady-state pCO 2 conditions in the modeled lung.
RESULTS: DCO2 trachea correlated strongly with pCO 2 , independent of the leak flow level (P < 0.0001). DCO 2 prong correlated with pCO 2 without and with moderate leak (P < 0.0001) but not with maximum leak (P = 0.1432). V T trachea correlated with the quotient of tracheal pressure amplitude and frequency irrespective of the leak (P < 0.0001). Based on the pressure amplitude at prong level (A prong ) V T trachea continued to follow a linear model of which the slopes decreased with increasing leak flow. V T prong correlated with the quotient of A prong and frequency, irrespective of the leak (P < 0.0001).
CONCLUSIONS: DCO2 obtained at the airway opening at prong level reflects ventilation efficacy during nHFOV even in the presence of moderate oropharyngeal leak.