An in-vitro study to evaluate high-volume low-pressure endotracheal tube cuff deflation dynamics.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Marti Joan D, Li Bassi Gianluigi, Isetta Valentina, Lazaro Miguel R, Aguilera-Xiol Eli, Comaru Talitha, Battaglini Denise, Meli Andrea, Ferrer Miguel, Navajas Daniel, Pelosi Paolo, Chiumello Davide, Torres Antoni, Farre Ramon
Аннотация
BACKGROUND: High-volume low-pressure (HVLP) endotracheal tube (ETT) cuffs for critically ill patients often deflate during the course of mechanical ventilation. We performed an in-vitro study to comprehensively assess HVLP cuff deflation dynamics and potential preventive measures.
METHODS: We evaluated 24-hour deflation of seven HVLP cuffs of cylindrical or tapered shape, and made of polyvinylchloride or polyurethane. Experiments were performed within a thermostated chamber set at 37 °C. In the first stage of experiments, the cuff pilot balloon valve was not manipulated. The cuff internal pressure was assessed hourly for 24 hours, via a linear position sensor which monitored cuff deflation displacements. Then, we re-evaluated cuff deflation of the worst-performing ETT cuffs with the cuff pilot balloon valve sealed. Finally, we inflated ETT cuffs within an artificial trachea to evaluate deflation dynamics during mechanical ventilation.
RESULTS: Initial tests showed an exponential decrease in cuff internal pressure in five out of seven cuffs. Cuffs of cylindrical shape and made of polyurethane demonstrated the fastest deflation rates (P<0.050 vs. cuffs of conical shape and made of polyvinylchloride). When the cuff pilot balloon valve was not sealed, the internal cuff pressure deflation rate differed significantly among ETTs (P=0.005). Yet, upon sealing the cuff pilot balloon valve and during mechanical ventilation, cuff deflation rates decreased (P<0.050).
CONCLUSIONS: In controlled in-vitro settings, ETT cuffs consistently deflate over time, and the cuff pilot balloon valve plays a central role in this occurrence. Deflation rate decreases when cuffs are inflated within a plastic artificial tracheal model and mechanical ventilation is activated.
METHODS: We evaluated 24-hour deflation of seven HVLP cuffs of cylindrical or tapered shape, and made of polyvinylchloride or polyurethane. Experiments were performed within a thermostated chamber set at 37 °C. In the first stage of experiments, the cuff pilot balloon valve was not manipulated. The cuff internal pressure was assessed hourly for 24 hours, via a linear position sensor which monitored cuff deflation displacements. Then, we re-evaluated cuff deflation of the worst-performing ETT cuffs with the cuff pilot balloon valve sealed. Finally, we inflated ETT cuffs within an artificial trachea to evaluate deflation dynamics during mechanical ventilation.
RESULTS: Initial tests showed an exponential decrease in cuff internal pressure in five out of seven cuffs. Cuffs of cylindrical shape and made of polyurethane demonstrated the fastest deflation rates (P<0.050 vs. cuffs of conical shape and made of polyvinylchloride). When the cuff pilot balloon valve was not sealed, the internal cuff pressure deflation rate differed significantly among ETTs (P=0.005). Yet, upon sealing the cuff pilot balloon valve and during mechanical ventilation, cuff deflation rates decreased (P<0.050).
CONCLUSIONS: In controlled in-vitro settings, ETT cuffs consistently deflate over time, and the cuff pilot balloon valve plays a central role in this occurrence. Deflation rate decreases when cuffs are inflated within a plastic artificial tracheal model and mechanical ventilation is activated.