Novel Split Chest Tube Improves Post-Surgical Thoracic Drainage.

Olivencia-Yurvati Albert H, Cherry Brandon H, Gurji Hunaid A, White Daniel W, Newton J Tyler, Scott Gary F, Hoxha Besim, Gourlay Terence, Mallet Robert T

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Conventional, separate mediastinal and pleural tubes are often inefficient at draining thoracic effusions.

DESCRIPTION: We developed a Y-shaped chest tube with split ends that divide within the thoracic cavity, permitting separate intrathoracic placement and requiring a single exit port. In this study, thoracic drainage by the split drain vs. that of separate drains was tested.

METHODS

After sternotomy, pericardiotomy, and left pleurotomy, pigs were fitted with separate chest drains (n=10) or a split tube prototype (n=9) with internal openings positioned in the mediastinum and in the costo-diaphragmatic recess. Separate series of experiments were conducted to test drainage of D5W or 0.58 M sucrose, an aqueous solution with viscosity approximating that of plasma. One litre of fluid was infused into the thorax, and suction was applied at -20 cm H2O for 30 min.

RESULTS

When D5W was infused, the split drain left a residual volume of 53 ± 99 ml (mean value ± SD) vs. 148 ± 120 for the separate drain (P=0.007), representing a drainage efficiency (i.e. drained vol/[drained + residual vol]) of 95 ± 10% vs. 86 ± 12% for the separate drains (P = 0.011). In the second series, the split drain evacuated more 0.58 M sucrose in the first minute (967 ± 129 ml) than the separate drains (680 ± 192 ml, P<0.001). By 30 min, the split drain evacuated a similar volume of sucrose vs. the conventional drain (1089 ± 72 vs. 1056 ± 78 ml; P = 0.5). Residual volume tended to be lower (25 ± 10 vs. 62 ± 72 ml; P = 0.128) and drainage efficiency tended to be higher (98 ± 1 vs. 95 ± 6%; P = 0.111) with the split drain vs. conventional separate drains.

CONCLUSION

The split chest tube drained the thoracic cavity at least as effectively as conventional separate tubes. This new device could potentially alleviate postoperative complications.

Для цитирования: Olivencia-Yurvati Albert H, Cherry Brandon H, Gurji Hunaid A, White Daniel W, Newton J Tyler, Scott Gary F, Hoxha Besim, Gourlay Terence, Mallet Robert T. Novel Split Chest Tube Improves Post-Surgical Thoracic Drainage.. Journal of clinical & experimental cardiology. 2014. DOI: 10.4172/2155-9880.1000321.