Evaluation of image quality of DWIBS versus DWI sequences in thoracic MRI at 3T.

Mesmann Caroline, Sigovan Monica, Berner Lise-Prune, Abergel Adva, Tronc François, Berthezène Y, Douek P, Boussel Loic

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare diffusion weighted imaging with background suppression (DWIBS) sequence with classic spectral diffusion sequence (DWI) with and without respiratory gating in mediastinal lymph node analysis at 3T.

MATERIALS AND

METHODS

26 patients scheduled for mediastinoscopic lymph node analysis, prospectively undergone a thoracic 3T MRI with DWIBS (FatSat=STIR; TR/TE=6674.1/44.7ms; IR=260 ms) and DWI sequences (FatSat=SPIR; TR/TE=1291/59.6 ms) (b=0-400-800 s/mm2) with and without (free breathing) respiratory gating. Images at b=800 were analyzed by two radiologists. They performed qualitative analysis of fat-sat homogeneity and motion artifacts, rated from 0 to 4, and quantitative evaluation by studying signal to background (STB) of lymph nodes.

RESULTS

Quality of fat suppression was significantly higher for DWIBS than for DWI both for free-breathing (score 3.48±0.65 vs. 1.76±0.96, p<0.0001) and respiratory-gated scans (3.17±0.77 vs. 1.72±0.73, p=0.0001). Similarly, artifacts were reduced with DWIBS (3.16±0.47 vs. 1.76±0.59, p<0.0001; 3.0±0.73 vs. 2.04±0.53, p=0.0001). Quantitative analysis showed higher STB with DWIBS (3.26±1.83 vs. 0.98±0.44, p<0.0001; 3.56±, 2.09 vs. 0.92±0.59, p<0.0001). Gating did not improve image quality and STB on DWIBS (p>0.05).

CONCLUSION

In thoracic MRI, ungated DWIBS sequence improves fat-sat homogeneity, reduces motion artifacts and increases STB of lymph nodes. Respiratory gating does not improve DWIBS image quality.

Для цитирования: Mesmann Caroline, Sigovan Monica, Berner Lise-Prune, Abergel Adva, Tronc François, Berthezène Y, Douek P, Boussel Loic. Evaluation of image quality of DWIBS versus DWI sequences in thoracic MRI at 3T.. Magnetic resonance imaging. 2014. DOI: 10.1016/j.mri.2014.08.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.