Journal of surgical oncology · 2021
Magnetic resonance imaging system for intraoperative margin assessment for DCIS and invasive breast cancer using the ClearSight™ system in breast-conserving surgery-Results from a postmarketing study.
Клиническое исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Thill Marc, Szwarcfiter Iris, Kelling Katharina, van Haasteren Viviane, Kolka Eyal, Noelke Josefa, Peles Zachi, Papa Moshe, Aulmann Sebastian, Allweis Tanir
Аннотация
BACKGROUND AND OBJECTIVES: Breast-conserving surgery (BCS) is followed by reoperations in approximately 25%. Reoperations lead to an increased risk of infection and wound healing problems as well as a worse cosmetic outcome. Several technical approaches for an intraoperative margin assessment to decrease the reoperation rate are under evaluation, some of them are still experimental.
METHODS: A prospective single-arm post-marketing study with 60 patients undergoing BCS for ductal carcinoma in situ (DCIS) and invasive breast cancer was conducted. The specimen was intraoperatively examined by the ClearSight™ system, a mobile magnetic resonance imaging system that is based on a diffusion-weighted imaging protocol. However, the results were blinded to the surgeon.
RESULTS: The ClearSight™ system was performed for both ductal and lobular breast cancer and DCIS, with a sensitivity of 0.80 (95% confidence interval [CI]: 0.44-0.96) and a specificity of 0.84 (95% CI 0.72-0.92), with an overall diagnostic accuracy of 80%.
CONCLUSION: Had the ClearSight™ been known to the surgeon intraoperatively, the reoperation rate would have been reduced by 83% for invasive carcinoma, from 10% to 2%, and 50% for DCIS, from 30% to 15% reoperations. A trial designed to examine the impact on reoperation rates is currently ongoing.
METHODS: A prospective single-arm post-marketing study with 60 patients undergoing BCS for ductal carcinoma in situ (DCIS) and invasive breast cancer was conducted. The specimen was intraoperatively examined by the ClearSight™ system, a mobile magnetic resonance imaging system that is based on a diffusion-weighted imaging protocol. However, the results were blinded to the surgeon.
RESULTS: The ClearSight™ system was performed for both ductal and lobular breast cancer and DCIS, with a sensitivity of 0.80 (95% confidence interval [CI]: 0.44-0.96) and a specificity of 0.84 (95% CI 0.72-0.92), with an overall diagnostic accuracy of 80%.
CONCLUSION: Had the ClearSight™ been known to the surgeon intraoperatively, the reoperation rate would have been reduced by 83% for invasive carcinoma, from 10% to 2%, and 50% for DCIS, from 30% to 15% reoperations. A trial designed to examine the impact on reoperation rates is currently ongoing.