MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/jso.26721

Медицинская библиотека · Journal of surgical oncology · 2021

Magnetic resonance imaging system for intraoperative margin assessment for DCIS and invasive breast cancer using the ClearSight™ system in breast-conserving surgery-Results from a postmarketing study.

Thill Marc, Szwarcfiter Iris, Kelling Katharina, van Haasteren Viviane, Kolka Eyal, Noelke Josefa, Peles Zachi, Papa Moshe, Aulmann Sebastian, Allweis Tanir

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVES

Breast-conserving surgery (BCS) is followed by reoperations in approximately 25%. Reoperations lead to an increased risk of infection and wound healing problems as well as a worse cosmetic outcome. Several technical approaches for an intraoperative margin assessment to decrease the reoperation rate are under evaluation, some of them are still experimental.

METHODS

A prospective single-arm post-marketing study with 60 patients undergoing BCS for ductal carcinoma in situ (DCIS) and invasive breast cancer was conducted. The specimen was intraoperatively examined by the ClearSight™ system, a mobile magnetic resonance imaging system that is based on a diffusion-weighted imaging protocol. However, the results were blinded to the surgeon.

RESULTS

The ClearSight™ system was performed for both ductal and lobular breast cancer and DCIS, with a sensitivity of 0.80 (95% confidence interval [CI]: 0.44-0.96) and a specificity of 0.84 (95% CI 0.72-0.92), with an overall diagnostic accuracy of 80%.

CONCLUSION

Had the ClearSight™ been known to the surgeon intraoperatively, the reoperation rate would have been reduced by 83% for invasive carcinoma, from 10% to 2%, and 50% for DCIS, from 30% to 15% reoperations. A trial designed to examine the impact on reoperation rates is currently ongoing.

Для цитирования: Thill Marc, Szwarcfiter Iris, Kelling Katharina, van Haasteren Viviane, Kolka Eyal, Noelke Josefa, Peles Zachi, Papa Moshe, Aulmann Sebastian, Allweis Tanir. Magnetic resonance imaging system for intraoperative margin assessment for DCIS and invasive breast cancer using the ClearSight™ system in breast-conserving surgery-Results from a postmarketing study.. Journal of surgical oncology. 2021. DOI: 10.1002/jso.26721.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.