Assessing the reproducibility of high definition urethral pressure profilometry and its correlation with an air-charged system.

Klünder Mario, Amend Bastian, Sawodny Oliver, Stenzl Arnulf, Ederer Michael, Kelp Alexandra, Sievert Karl-Dietrich, Feuer Ronny

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Recently, a new urodynamic method for the assessment of stress urinary incontinence called high definition urethral pressure profilometry (HD-UPP) has been introduced. This method combines a novel microtip catheter with advanced signal processing to enable spatial data location and the reconstruction of a pressure image inside the urethra. In order to assess the reproducibility of HD-UPP data, we statistically evaluate HD-UPP datasets and compare them to data from a double balloon air-charged system.

MATERIALS AND

METHODS

Both catheters are used on sedated female minipigs. Data from the microtip catheter are processed through a signal reconstruction algorithm, urodynamic features are extracted, and compared to the air-charged system. Reproducibility of HD-UPP data is assessed by statistically evaluating consecutive, intra-individual datasets.

RESULTS

HD-UPP delivers results in agreement with previous comparisons of microtip and air-charged systems. The average deviation of two consecutive, intra-individual pressure images is very low at 7 cm H2 O.

CONCLUSIONS

HD-UPP provides physicians with detailed information on the pressure distribution inside the urethra. Through comparison with an air-charged catheter, it is shown that HD-UPP delivers results in agreement with previous studies on the comparison of microtip and air-charged catheters. It provides excellent reproducibility, as the difference between sequentially measured profiles from the same minipig is significantly lower than the one between profiles from different minipigs.

Для цитирования: Klünder Mario, Amend Bastian, Sawodny Oliver, Stenzl Arnulf, Ederer Michael, Kelp Alexandra, Sievert Karl-Dietrich, Feuer Ronny. Assessing the reproducibility of high definition urethral pressure profilometry and its correlation with an air-charged system.. Neurourology and urodynamics. 2016. DOI: 10.1002/nau.23115.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.