International ophthalmology · 2023
Atomic force microscopy for the evaluation of corneal surface roughness after femtosecond laser flap creation and excimer ablation.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Cañones-Zafra Rafael, Gros-Otero Juan, Garcia-Gonzalez Montserrat, Casado Santiago, Ketabi Samira, Villa-Collar César, Teus Miguel A
Аннотация
INTRODUCTION: It is well known that the femtosecond laser lamellar cut induces some degree of surface roughness. Nevertheless, as in femtosecond laser-assisted LASIK (FS-LASIK), an excimer LASIK ablation is performed, and the post-ablation stromal bed should show some degree of smoothening. We decided to compare, using atomic force microscopy (AFM), the roughness of the corneal stromal bed, after a femtosecond lasers device flap was created with or without an excimer myopic ablation.
METHODS: Using 6 freshly enucleated porcine eyes, we created in every eye a flap using a femtosecond laser. Additionally, in 3 eyes, an excimer laser ablation to correct-3 diopters (D) was made. AFM imaging of the remaining corneal stroma was performed. Ten different square areas of 20 μm x 20 μm at the central area of the stroma of each corneal sample were studied. The roughness parameters used were the root-mean-square deviation from a perfectly flat surface.
RESULTS: The RMS deviation was 360 ± 120 nm in femtosecond laser only, and 110 ± 20 nm in those cases where excimer is also involved (p < 0.0001).
CONCLUSIONS: Our results show that the roughness of the surface treated with excimer is clearly lower than in the group with no excimer ablation; thus, the application of laser excimer after a flap created by femtosecond laser seems to soften the nano-irregularities created by this technique.
METHODS: Using 6 freshly enucleated porcine eyes, we created in every eye a flap using a femtosecond laser. Additionally, in 3 eyes, an excimer laser ablation to correct-3 diopters (D) was made. AFM imaging of the remaining corneal stroma was performed. Ten different square areas of 20 μm x 20 μm at the central area of the stroma of each corneal sample were studied. The roughness parameters used were the root-mean-square deviation from a perfectly flat surface.
RESULTS: The RMS deviation was 360 ± 120 nm in femtosecond laser only, and 110 ± 20 nm in those cases where excimer is also involved (p < 0.0001).
CONCLUSIONS: Our results show that the roughness of the surface treated with excimer is clearly lower than in the group with no excimer ablation; thus, the application of laser excimer after a flap created by femtosecond laser seems to soften the nano-irregularities created by this technique.