Comprehensive dosimetric comparison of halcyon 3.0 versus truebeam for breast cancer volumetric modulated arc therapy plans.

Gu Anyan, Wang Mengyuan, Pan Zhenyu, Tang Ran, Li Yaotao, Wu Yu, Yang Guozi, Sun Xingru

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Breast cancer is the foremost malignancy threatening female health. This study aimed to compare the dosimetric performance of Halcyon 3.0 and TrueBeam in Volumetric Modulated Arc Therapy (VMAT) planning for breast cancer.

METHODS

Thirty post-operative breast cancer patients were included. VMAT plans for Halcyon 3.0 and TrueBeam were generated using 6MV FFF beams in Eclipse 16.1 with identical optimizing objectives. Organs At Risk (OARs) and target metrics were evaluated. Statistical analysis employed paired t-tests or Wilcoxon signed-rank tests (α=0.05).

RESULTS

Halcyon significantly reduced the mean dose of heart (Δ = -112.5 cGy, P< 0.001), left anterior descending artery mean dose (Δ = -151.3 cGy, P< 0.001), and spinal cord maximum dose (Δ = -88.2 cGy, P = 0.011). Low-dose exposure improved, with heart V5 (volume receiving ≥500 cGy) reduced from 22.0% to 12.6% (P< 0.001) and ipsilateral lung V5 reduced from 59.4% to 45.8% (P< 0.001) between Halcyon and TrueBeam. Halcyon lowered the absolute volume received 500 cGy by 487 cm³ (P< 0.001). Halcyon had the better target homogeneity (homogeneity index: Δ = -0.01, P< 0.001), while conformity remained comparable (P > 0.05). Despite having higher monitor units (Δ = 87.5, P< 0.001), Halcyon maintained delivery efficiency through faster leaf motion speed and gantry rotation velocity.

CONCLUSIONS

Compared with TrueBeam, Halcyon achieved reduction in low-dose exposure while maintaining target coverage in breast cancer VMAT planning, thereby lowering radiation doses to OARs (e.g., heart and lung V5). This benefit is likely attributed to Halcyon's unique staggered dual-layer MLC design.

Для цитирования: Gu Anyan, Wang Mengyuan, Pan Zhenyu, Tang Ran, Li Yaotao, Wu Yu, Yang Guozi, Sun Xingru. Comprehensive dosimetric comparison of halcyon 3.0 versus truebeam for breast cancer volumetric modulated arc therapy plans.. Frontiers in oncology. 2025. DOI: 10.3389/fonc.2025.1630702.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.