Canadian Registry of Implantable Electronic Device Outcomes: Surveillance of the Riata Lead Under Advisory.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Parkash Ratika, Thibault Bernard, Mangat Iqwal, Coutu Benoit, Bennett Matthew, Healey Jeffrey, Verma Atul, Philippon Francois, Sandhu Roopinder, Cameron Douglas, Ayala-Paredes Felix, Sterns Laurence, Essebag Vidal, Kus Teresa, Nery Pablo, Stephenson Elizabeth, Yee Raymond, Exner Derek, Toal Satish, Birnie David, Wells George, Krahn Andrew
Аннотация
BACKGROUND: The Riata lead under advisory has posed a unique clinical scenario where inside-out abrasion results in externalization of conductor cables, with a higher risk of electrical failure. We developed a comprehensive registry to assist with clinical management of this lead.
METHODS AND RESULTS: This Canadian registry reports the findings of 3763 (74.2% of all Riata leads in Canada) Riata leads under advisory, with a mean follow-up time of 8.9±1.5 years. The overall electrical failure rate was 5.2% at 8 years, with no difference between 7-French and 8-French lead models. Cable externalization was found to be more common in the 8-French model (12.3% versus 5.2%, P<0.0001) and was associated with a higher risk of electrical failure. Predictors of electrical lead failure included cable externalization, higher left ventricular ejection fraction, younger age, higher body mass index, and a passive fixation lead. One patient died due to electrical failure, a further 2 patients survived an event where the device failed to deliver high-voltage therapy. Major complications because of lead extraction were higher when compared with lead abandonment, no difference among lead model observed. Two deaths occurred as a consequence of lead extraction, in the context of an underlying infection.
CONCLUSIONS: The Riata lead under advisory has a steady electrical failure rate over time. There are identifiable predictors of lead failure that can assist with clinical decisions as to whether lead revision should be performed prophylactically.
METHODS AND RESULTS: This Canadian registry reports the findings of 3763 (74.2% of all Riata leads in Canada) Riata leads under advisory, with a mean follow-up time of 8.9±1.5 years. The overall electrical failure rate was 5.2% at 8 years, with no difference between 7-French and 8-French lead models. Cable externalization was found to be more common in the 8-French model (12.3% versus 5.2%, P<0.0001) and was associated with a higher risk of electrical failure. Predictors of electrical lead failure included cable externalization, higher left ventricular ejection fraction, younger age, higher body mass index, and a passive fixation lead. One patient died due to electrical failure, a further 2 patients survived an event where the device failed to deliver high-voltage therapy. Major complications because of lead extraction were higher when compared with lead abandonment, no difference among lead model observed. Two deaths occurred as a consequence of lead extraction, in the context of an underlying infection.
CONCLUSIONS: The Riata lead under advisory has a steady electrical failure rate over time. There are identifiable predictors of lead failure that can assist with clinical decisions as to whether lead revision should be performed prophylactically.