MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1213/xaa.0000000000000404

Медицинская библиотека · A & A case reports · 2017

Difficulty Inserting Cuffed Endotracheal Tubes in a Child: A Case Report.

Imai Keiko, Doi Taku, Kayashima Kenji

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We experienced difficulty inserting cuffed inner diameter (ID) 4.5- and 5.0-mm endotracheal tubes (ETTs) in a 5-year-old boy. Postoperative ultrasound investigations showed that the internal transverse width of the cricoid cartilage was 8.0 mm. The maximum outer diameter (OD) of the deflated cuff portion of the cuffed ID 4.5- and 5.0-mm ETTs was 8.5 and 9.6 mm, respectively. The OD of an uncuffed ID 5.5-mm ETT was 7.6 mm; this tube passed the cricoid cartilage. Hence, the transverse width of the cricoid cartilage and ETT diameter including cuff folds should be considered when selecting cuffed ETTs.

Для цитирования: Imai Keiko, Doi Taku, Kayashima Kenji. Difficulty Inserting Cuffed Endotracheal Tubes in a Child: A Case Report.. A & A case reports. 2017. DOI: 10.1213/xaa.0000000000000404.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.