DOI 10.1002/1097-0320(20010101)43:1<2::aid-cyto1012>3.0.co;2-j

Медицинская библиотека · Cytometry · 2001

NASA/American Cancer Society High-Resolution Flow Cytometry Project-I.

Thomas R A, Krishan A, Robinson D M, Sams C, Costa F, New Collective Author

СтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The NASA/American Cancer Society (ACS) flow cytometer can simultaneously analyze the electronic nuclear volume (ENV) and DNA content of cells. This study describes the schematics, resolution, reproducibility, and sensitivity of biological standards analyzed on this unit.

METHODS

Calibrated beads and biological standards (lymphocytes, trout erythrocytes [TRBC], calf thymocytes, and tumor cells) were analyzed for ENV versus DNA content. Parallel data (forward scatter versus DNA) from a conventional flow cytometer were obtained.

RESULTS

ENV linearity studies yielded an R value of 0.999. TRBC had a coefficient of variation (CV) of 1.18 +/- 0.13. DNA indexes as low as 1.02 were detectable. DNA content of lymphocytes from 42 females was 1.9% greater than that for 60 males, with a noninstrumental variability in total DNA content of 0.5%. The ENV/DNA ratio was constant in 15 normal human tissue samples, but differed in the four animal species tested. The ENV/DNA ratio for a hypodiploid breast carcinoma was 2.3 times greater than that for normal breast tissue.

CONCLUSIONS

The high-resolution ENV versus DNA analyses are highly reliable, sensitive, and can be used for the detection of near-diploid tumor cells that are difficult to identify with conventional cytometers. ENV/DNA ratio may be a useful parameter for detection of aneuploid populations.

Для цитирования: Thomas R A, Krishan A, Robinson D M, Sams C, Costa F, New Collective Author. NASA/American Cancer Society High-Resolution Flow Cytometry Project-I.. Cytometry. 2001. DOI: 10.1002/1097-0320(20010101)43:1<2::aid-cyto1012>3.0.co;2-j.