DOI 10.1186/s12880-024-01228-1

Медицинская библиотека · BMC medical imaging · 2024

Influence of spectral shaping and tube voltage modulation in ultralow-dose computed tomography of the abdomen.

Feldle Philipp, Grunz Jan-Peter, Kunz Andreas Steven, Pannenbecker Pauline, Patzer Theresa Sophie, Pichlmeier Svenja, Sauer Stephanie Tina, Hendel Robin, Ergün Süleyman, Bley Thorsten Alexander, Huflage Henner

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Unenhanced abdominal CT constitutes the diagnostic standard of care in suspected urolithiasis. Aiming to identify potential for radiation dose reduction in this frequent imaging task, this experimental study compares the effect of spectral shaping and tube voltage modulation on image quality.

METHODS

Using a third-generation dual-source CT, eight cadaveric specimens were scanned with varying tube voltage settings with and without tin filter application (Sn 150, Sn 100, 120, 100, and 80 kVp) at three dose levels (3 mGy: standard; 1 mGy: low; 0.5 mGy: ultralow). Image quality was assessed quantitatively by calculation of signal-to-noise ratios (SNR) for various tissues (spleen, kidney, trabecular bone, fat) and subjectively by three independent radiologists based on a seven-point rating scale (7 = excellent; 1 = very poor).

RESULTS

Irrespective of dose level, Sn 100 kVp resulted in the highest SNR of all tube voltage settings. In direct comparison to Sn 150 kVp, superior SNR was ascertained for spleen (p ≤ 0.004) and kidney tissue (p ≤ 0.009). In ultralow-dose scans, subjective image quality of Sn 100 kVp (median score 3; interquartile range 3-3) was higher compared with conventional imaging at 120 kVp (2; 2-2), 100 kVp (1; 1-2), and 80 kVp (1; 1-1) (all p < 0.001). Indicated by an intraclass correlation coefficient of 0.945 (95% confidence interval: 0.927-0.960), interrater reliability was excellent.

CONCLUSIONS

In abdominal CT with maximised dose reduction, tin prefiltration at 100 kVp allows for superior image quality over Sn 150 kVp and conventional imaging without spectral shaping.

Для цитирования: Feldle Philipp, Grunz Jan-Peter, Kunz Andreas Steven, Pannenbecker Pauline, Patzer Theresa Sophie, Pichlmeier Svenja, Sauer Stephanie Tina, Hendel Robin, Ergün Süleyman, Bley Thorsten Alexander, Huflage Henner. Influence of spectral shaping and tube voltage modulation in ultralow-dose computed tomography of the abdomen.. BMC medical imaging. 2024. DOI: 10.1186/s12880-024-01228-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.