Anterior chamber temperature rise during torsional versus longitudinal phacoemulsification: an experimental porcine eye study.

Suzuki Hisaharu

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Previous studies have reported lower thermal profiles with torsional ultrasound than with longitudinal ultrasound; however, the mechanical determinants underlying this difference remain incompletely understood. In particular, the mechanical determinants of intraocular temperature behavior under controlled experimental conditions remain unclear. Therefore, this study aimed to compare anterior chamber temperature behavior between longitudinal and torsional ultrasound under standardized experimental conditions, including settings with nominally comparable tip-excursion parameters.

METHODS

Anterior chamber temperature changes were evaluated using porcine eyes under standardized experimental conditions. Continuous longitudinal and torsional ultrasound were applied with controlled irrigation and aspiration settings. Ultrasound power settings were adjusted to create nominally comparable tip-excursion conditions between the two modalities. Temperature was continuously recorded, and temporal temperature profiles were analyzed.

RESULTS

Under identical machine settings, torsional ultrasound produced a significantly smaller increase in anterior chamber temperature than longitudinal ultrasound. However, when nominally comparable tip-excursion settings were used, temperature rise did not differ significantly between the two modalities. In both ultrasound modes, anterior chamber temperature increased rapidly after activation and reached a plateau within approximately 20 s.

CONCLUSIONS

Under controlled low-flow experimental conditions, torsional ultrasound produced a smaller increase in anterior chamber temperature than longitudinal ultrasound. However, this difference was not observed under nominally comparable machine-level tip-excursion settings. These findings suggest that tip-motion characteristics may contribute to intraocular temperature behavior during ultrasound activation.

Для цитирования: Suzuki Hisaharu. Anterior chamber temperature rise during torsional versus longitudinal phacoemulsification: an experimental porcine eye study.. BMC ophthalmology. 2026. DOI: 10.1186/s12886-026-04968-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.