DOI 10.1016/j.amjoto.2023.104153

Медицинская библиотека · American journal of otolaryngology · 2023

Upper airway stimulation device failure: A 7-year single center experience.

Alapati Rahul, Wagoner Sarah F, Nieves Antonio Bon, Lawrence Amelia, Rouse David, Larsen Christopher

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To identify and report a single center experience with upper airway stimulator device-related failures.

STUDY

DESIGN

Retrospective case series.

SETTING

Single tertiary academic center.

METHODS

Retrospective data on 352 patients who underwent UAS surgery with an Inspire device from 2016 to 2023 was collected, including demographics, comorbidities, and nature of device failure requiring revision surgery.

RESULTS

Out of the 348 patients included in our analysis, 16 (4.6 %) required revision due to device failure, with an average interval of 772 days (∼2 years) between initial implant and revision. Most failures were attributed to respiratory sensing lead damage (n = 11, 68.8 %), resulting in high system impedance and subsequent device malfunction. Lead fracture causes varied, including idiopathic occurrences and potential trauma. Lead migration was noted in one case (6.3 %), where the hypoglossal electrode detached from the nerve. Two patients (12.3 %) required implantable pulse generator (IPG) replacement, one after experiencing trauma and the other due to unclear source of malfunction. One patient (6.3 %) required complete system replacement following high lead impedance and absent tongue motion. The last patient required replacement of both the IPG and respiratory lead after experiencing high lead impedance (6.3 %).

CONCLUSION

Respiratory sensing lead fracture emerged as the leading cause of device failure in this cohort, underscoring the need to address this under-reported issue, potentially linked to the time lapse after device implantation.

Для цитирования: Alapati Rahul, Wagoner Sarah F, Nieves Antonio Bon, Lawrence Amelia, Rouse David, Larsen Christopher. Upper airway stimulation device failure: A 7-year single center experience.. American journal of otolaryngology. 2023. DOI: 10.1016/j.amjoto.2023.104153.