First clinical evaluation of a new long-term subconjunctival glucose sensor.
Аннотация
BACKGROUND
To evaluate the feasibility of an implantable subconjunctival glucose monitoring system (SGMS) for glucose monitoring in humans, we investigated the in vivo performance of the sensor in a clinical trial with five patients.
METHODS
The new SGMS consists of an implantable ocular mini implant (OMI) and a hand-held fluorescence photometer. The implantable subconjunctival glucose sensor is composed of a fluorescence resonance energy transfer system based on Concanavalin A chemistry, embedded in a nelfilcon polymer hydrogel disk. Blood glucose changes in humans were induced by oral glucose intake and insulin injections.
RESULTS
The in vivo response of the new SGMS was tested in a first human clinical study with five diabetes patients. The OMI was well tolerated in the eyes of the patients. The SGMS exhibited high correlation coefficients (>0.88) with blood glucose changes and a good stability of the sensor response to glucose for the study period of 2 weeks. Lag times were in the range of 5-10 min. A total of 98% of all data pairs was in the clinical acceptable ranges A and B of the consensus error grid.
CONCLUSIONS
For the first time, the possibility to measure glucose in vivo in the subconjunctival interstitial fluid for a period of 2 weeks was demonstrated in a human clinical trial.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.