Estimation of skin sodium content via 23Na magnetic resonance imaging at 3 T using a single-loop surface transmit/receive coil.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Birchall Jonathan Richard, Kaggie Joshua Daniel, Riemer Frank, Selvarajah Viknesh, Connolly Kathleen, Graves Martin John, Wilkinson Ian, Gallagher Ferdia Aiden, McLean Mary Anne
Аннотация
OBJECTIVE: To assess the feasibility of performing quantitative in vivo tissue sodium concentration measurements in the skin with MRI using a sodium-tuned single-loop transmit/receive surface coil at 3 T.
MATERIALS AND METHODS: Proton and sodium imaging of skin at the lower back was performed in 17 healthy volunteers on a clinical 3 T system using an anisotropic (1 x 1 x 10 mm) 3D cones sequence. B1 mapping via the double-angle method was used to account for spatial variation in transmit and receive sensitivity. Apparent tissue sodium concentration (aTSC) was estimated relative to a concurrently imaged phantom of known concentration.
RESULTS: The mean and standard deviation of B1-corrected aTSC across all subjects were 28 ± 7 mM, approximately 63% higher than previously published studies at 3 T. Although the study was not originally powered to detect gender or racial differences, a higher aTSC was observed in males relative to females (31 ± 5 mM vs. 23 ± 7 mM, distribution 11M/6F).
DISCUSSION: The skin aTSC values recorded were slightly higher than in some previous studies, which possibly arose from the higher spatial resolution and coil sensitivity utilised here. The recorded values remain lower than those measured at 7 T or via direct tissue sampling. Sex-linked differences in 23Na-MRI-derived aTSC are consistent with biopsy results described in the literature. A scan duration of ~40 minutes was considered clinically acceptable.
MATERIALS AND METHODS: Proton and sodium imaging of skin at the lower back was performed in 17 healthy volunteers on a clinical 3 T system using an anisotropic (1 x 1 x 10 mm) 3D cones sequence. B1 mapping via the double-angle method was used to account for spatial variation in transmit and receive sensitivity. Apparent tissue sodium concentration (aTSC) was estimated relative to a concurrently imaged phantom of known concentration.
RESULTS: The mean and standard deviation of B1-corrected aTSC across all subjects were 28 ± 7 mM, approximately 63% higher than previously published studies at 3 T. Although the study was not originally powered to detect gender or racial differences, a higher aTSC was observed in males relative to females (31 ± 5 mM vs. 23 ± 7 mM, distribution 11M/6F).
DISCUSSION: The skin aTSC values recorded were slightly higher than in some previous studies, which possibly arose from the higher spatial resolution and coil sensitivity utilised here. The recorded values remain lower than those measured at 7 T or via direct tissue sampling. Sex-linked differences in 23Na-MRI-derived aTSC are consistent with biopsy results described in the literature. A scan duration of ~40 minutes was considered clinically acceptable.