Pleural fluid: Are temperature and storage time critical preanalytical error factors in biochemical analyses?
Аннотация
BACKGROUND
Biochemical analysis of fluid is the primary laboratory approach in pleural effusion diagnosis. Standardization of the steps between collection and laboratorial analyses are fundamental to maintain the quality of the results. We evaluated the influence of temperature and storage time on sample stability.
METHODS
Pleural fluid from 30 patients was submitted to analyses of proteins, albumin, lactic dehydrogenase (LDH), cholesterol, triglycerides, and glucose. Aliquots were stored at 21 degrees , 4 degrees , and-20 degrees C, and concentrations were determined after 1, 2, 3, 4, 7, and 14 days. LDH isoenzymes were quantified in 7 random samples.
RESULTS
Due to the instability of isoenzymes 4 and 5, a decrease in LDH was observed in the first 24h in samples maintained at -20 degrees C and after 2 days when maintained at 4 degrees C. Aside from glucose, all parameters were stable for up to at least day 4 when stored at room temperature or 4 degrees C.
CONCLUSIONS
Temperature and storage time are potential preanalytical errors in pleural fluid analyses, mainly if we consider the instability of glucose and LDH. The ideal procedure is to execute all the tests immediately after collection. However, most of the tests can be done in refrigerated samples, excepting LDH analysis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.