Evaluation of the detection ability of uropathogen morphology and vaginal contamination by the Atellica UAS800 automated urine microscopy analyzer and its effectiveness.

Nakamura Akihiro, Shinke Tetsuya, Noguchi Nobuyoshi, Komatsu Masaru, Yamanishi Hachiro

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

To help combat the worldwide spread of multidrug-resistant Enterobacterales, which are responsible for many causes of urinary tract infection (UTI), we evaluated the ability of the Atellica UAS800 automated microscopy system, the only one offering the capability of bacterial morphological differentiation, to determine its effectiveness.

METHODS

We examined 118 outpatient spot urine samples in which pyuria and bacteriuria were observed using flow cytometry (training set: 81; cross-validation set: 37). The ability of the Atellica UAS800 to differentiate between bacilli and cocci was verified. To improve its ability, multiple logistic regression analysis was used to construct a prediction formula.

RESULTS

This instrument's detection sensitivity was 106 CFU/ml, and reproducibility in that range was good, but data reliability for the number of cocci was low. Multiple logistic regression analysis with each explanatory variable (14 items from the Atellica UAS800, age and sex) showed the best prediction formula for discrimination of uropathogen morphology was a model with 5 explanatory variables: number of bacilli (p < 0.001), squamous epithelial cells (p = 0.004), age (p = 0.039), number of cocci (p = 0.107), and erythrocytes (p = 0.111). For a predicted cutoff value of 0.449, sensitivity was 0.879 and specificity was 0.854. In the cross-validation set, sensitivity was 0.813 and specificity was 0.857.

CONCLUSIONS

The Atellica UAS800 could detect squamous epithelial cells, an indicator of vaginal contamination, with high sensitivity, which further improved performance. Simultaneous use of this probability prediction formula with urinalysis results may facilitate real-time prediction of uropathogens and vaginal contamination, thus providing helpful information for empiric therapy.

Для цитирования: Nakamura Akihiro, Shinke Tetsuya, Noguchi Nobuyoshi, Komatsu Masaru, Yamanishi Hachiro. Evaluation of the detection ability of uropathogen morphology and vaginal contamination by the Atellica UAS800 automated urine microscopy analyzer and its effectiveness.. Journal of clinical laboratory analysis. 2021. DOI: 10.1002/jcla.23698.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.