Safety of heat generated by endoscope light sources in simulated transcanal endoscopic ear surgery.

Ito Tsukasa, Kubota Toshinori, Takagi Akira, Watanabe Tomoo, Futai Kazunori, Furukawa Takatoshi, Kakehata Seiji

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine whether heat generated by endoscope light sources during ear surgery is safe.

METHODS

Transcanal endoscopic ear surgery (TEES) was simulated using 2.7-mm or 4-mm endoscopes coupled to xenon or LED light sources and a 3D model of human temporal bone. The endoscope tip was fixed at the center of tympanic annulus. Light sources were tested at clinical (30% for xenon and 40% for LED) and 100% settings. Temperatures were measured using thermocouples attached to the endoscope tip and three points within the middle ear cavity: promontory, horizontal portion of the facial nerve and lateral semicircular canal.

RESULTS

Maximum temperatures measured within the middle ear cavity were below 31°C at clinical settings, while the temperatures rose to 44.1°C using a 4-mm endoscope with a xenon light source set at 100%. Temperatures measured at the tip were all safe at clinical settings, but rose dramatically to 110.1°C for the 4-mm endoscope with xenon at 100%.

CONCLUSION

Endoscopes can be safely used within the middle ear at clinical settings. However, operators should not exceed clinical settings, particularly with 4-mm endoscopes with a xenon light source, to ensure temperatures generated within the middle ear cavity are safe.

Для цитирования: Ito Tsukasa, Kubota Toshinori, Takagi Akira, Watanabe Tomoo, Futai Kazunori, Furukawa Takatoshi, Kakehata Seiji. Safety of heat generated by endoscope light sources in simulated transcanal endoscopic ear surgery.. Auris, nasus, larynx. 2016. DOI: 10.1016/j.anl.2015.12.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.