DOI 10.3928/15428877-20100929-06

Медицинская библиотека · Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye · 2010

Narrow- and open-angle measurements with anterior-segment optical coherence tomography and Pentacam™.

Mou Dapeng, Fu Jing, Li Shuning, Wang Lan, Wang Xiaozhen, Wu Gewei, Qing Guoping, Peng Yi, Wang Ningli

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

To assess the agreement of the Pentacam (Oculus, Wetzlar, Germany) and anterior-segment optical coherence tomography (AS-OCT) in measuring open and narrow angles.

PATIENTS AND

METHODS

Prospective observational case series in which 39 healthy normal subjects (39 eyes) and 25 patients with narrow angles (37 eyes) were enrolled between May and September 2008. The anterior chamber measurements were performed by AS-OCT and the Pentacam.

RESULTS

There was no significant difference in the open-angle measurements between the Pentacam and AS-OCT (P > .05). However, in patients with narrow angles, the average values of nasal and temporal anterior chamber angles taken by the Pentacam were larger than those of AS-OCT (25.5° ± 5.66° and 25.77° ± 5.15° vs 13.40° ± 6.81° and 12.13° ± 6.47°; P < .001).

CONCLUSION

Both AS-OCT and the Pentacam can reliably measure anterior chamber angles in healthy normal subjects. The difference in measuring narrow angles by the two instruments was due to inability to view the angle recess or the scleral spur with the Pentacam.

Для цитирования: Mou Dapeng, Fu Jing, Li Shuning, Wang Lan, Wang Xiaozhen, Wu Gewei, Qing Guoping, Peng Yi, Wang Ningli. Narrow- and open-angle measurements with anterior-segment optical coherence tomography and Pentacam™.. Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye. 2010. DOI: 10.3928/15428877-20100929-06.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.