Attenuation coefficient and sound speed in human myometrium and uterine fibroid tumors.
Аннотация
OBJECTIVE
To develop a noninvasive method for treatment of uterine fibroid tumors using high-intensity focused ultrasound. Optimal high-intensity focused ultrasound treatment would be dependent on quantitative information about ultrasonic tissue characteristics.
METHODS
Ultrasonic attenuation and the sound speed of fresh human fibroid tumors and myometrium were measured as a function of frequency (1-3 MHz) by using a pulse transmission technique before and after in vitro high-intensity focused ultrasound treatment (3.5 MHz at an intensity of 2,000 W/cm2).
RESULTS
The ranges of the attenuation coefficients, before and after high-intensity focused ultrasound treatment, were 0.9 to 2.2 and 1.8 to 3.9 dB/cm2, respectively, for fibroid tumors and 0.5 to 1.6 and 1.7 to 3.3 dB/cm2, respectively, for myometrium. Although the sound speed appeared to be independent of frequency (1,611 to 1,616 m/s at 1 to 3 MHz) in both types of tissues, a slight increase of approximately 4 to 14 m/s was observed after high-intensity focused ultrasound treatment.
CONCLUSIONS
The results of this study represent our first reported values of the attenuation coefficient and sound speed in fibroid tumors and myometrium before and after high-intensity focused ultrasound treatment.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.