An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients.
Аннотация
BACKGROUND
Although core temperature can be measured invasively, there are currently no widely available, reliable, noninvasive thermometers for its measurement. We thus compared a prototype zero-heat-flux thermometer with simultaneous measurements from a pulmonary artery catheter. Specifically, we tested the hypothesis that zero-heat-flux temperatures are sufficiently accurate for routine clinical use.
METHODS
Core temperature was measured from the thermistor of a standard pulmonary artery catheter and with a prototype zero-heat-flux deep-tissue thermometer in 105 patients having nonemergent cardiac surgery. Zero-heat-flux probes were positioned on the lateral forehead and lateral neck. Skin surface temperature probes were attached to the forehead just adjacent to the zero-heat-flux probe. Temperatures were recorded at 1-minute intervals, excluding the period of cardiopulmonary bypass, and for the first 4 postoperative hours. Zero-heat-flux and pulmonary artery temperatures were compared with bias analysis; differences exceeding 0.5°C were considered to be potentially clinically important.
RESULTS
The mean duration in the operating room was 279 ± 75 minutes, and the mean cross-clamp time was 118 ± 50 minutes. All subjects were monitored for an additional 4 hours in the intensive care unit. The average overall difference between forehead zero-heat-flux and pulmonary artery temperatures (i.e., forehead minus pulmonary artery) was -0.23°C (95% limits of agreement of ±0.82); 78% of the differences were ≤0.5°C. The average intraoperative temperature difference was -0.08°C (95% limits of agreement of ±0.88); 84% of the differences were ≤0.5°C. The average postoperative difference was -0.32°C (95% limits of agreement of ±0.75); 84% of the differences were ≤0.5°C. Bias and precision values for neck site were similar to the forehead values. Uncorrected forehead skin temperature showed an increasing negative bias as core temperature decreased.
CONCLUSIONS
Core temperature can be noninvasively measured using the zero-heat-flux method. Bias was small, but precision was slightly worse than our designated 0.5°C limits compared with measurements from a pulmonary artery catheter.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.