BACKGROUND
Atlantoaxial joint distraction has been advocated for the decompression of the brain stem in patients affected by basilar invagination, avoiding direct transoral decompression. This technique requires C2 ganglion resection and it is often impossible to perform due to the peculiar bony anatomy. We describe a cadaveric anatomical study supporting the feasibility of C1-C2 distraction performed with an expandable device, allowing easier insertion of the tool and preservation of the C2 nerve root.
METHODS
In five adult cadaveric specimens, posterior atlantoaxial surgical exposure was performed and an expandable system was inserted within the C1-C2 joint. The expansion of the device, leading to active distraction of the joint space, together with all the surgical steps of the technique was recorded with anatomical pictures and the final results were checked with a computed tomography (CT) scan.
RESULTS
Insertion of the device was easily performed in all cases without anatomical conflict with the C2 ganglion; CT scans confirmed the distraction of the C1-C2 joint.
CONCLUSION
This cadaveric anatomical study confirms the feasibility of the introduction of an expandable and flexible device within the C1-C2 joint, allowing it's distraction and preservation of the C2 ganglion.