DOI 10.1007/bf03014466

Медицинская библиотека · Canadian journal of anaesthesia = Journal canadien d'anesthesie · 1997

Comparison of distal oesophageal temperature with "deep" and tracheal temperatures.

Matsukawa T, Sessler D I, Ozaki M, Hanagata K, Iwashita H, Kumazawa T

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПоддержка службы здравоохранения СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare distal oesophageal (reference) temperature with "deep-sternal," "deep-forehead," and tracheal temperatures, establishing the accuracy and precision of each.

METHODS

We studied 20 patients undergoing general anaesthesia for gynaecological surgery. Their lungs were mechanically ventilated with a circle system, at a fresh-gas flow rate of 6 L.min-1 Respiratory gases were not warmed or humidified. Tracheal temperatures were recorded from a Trachelon tube inserted approximately 21 cm. Deep-body temperatures were measured at the sternum and forehead using a Coretemp thermometer. The principle of the method is to null thermal flux through a cutaneous disk, thus obliterating thermal gradients between the sides of the disk, skin surface, and subcutaneous tissues. Distal oesophageal temperatures were measured from thermocouples incorporated into oesophageal stethoscopes. Tracheal and deep-tissue temperatures were compared with oesophageal temperature using regression and Bland and Altman analyses.

RESULTS

Tracheal, sternal, and forehead temperatures correlated similarly with distal oesophageal temperature, correlation coefficients (r2) being 0.7 in each case. The offset (oesophageal temperature minus study site) was considerably larger for tracheal temperature (0.7 degree C) than for the other sites (0.2 degree C). However, the precision was only 0.3 degree C at each site.

CONCLUSION

Our data suggest that tracheal temperatures may not be an adequate substitute for conventional core-temperature monitoring sites. In contrast, the accuracy and precision of deep-tissue temperature monitoring at the sternum and forehead was sufficient for clinical use.

Для цитирования: Matsukawa T, Sessler D I, Ozaki M, Hanagata K, Iwashita H, Kumazawa T. Comparison of distal oesophageal temperature with "deep" and tracheal temperatures.. Canadian journal of anaesthesia = Journal canadien d'anesthesie. 1997. DOI: 10.1007/bf03014466.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.