Flow injection analysis vs. ultra high performance liquid chromatography coupled with tandem mass spectrometry for determination of imatinib in human plasma.

Mičová Kateřina, Friedecký David, Faber Edgar, Polýnková Adriana, Adam Tomáš

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The aim of this study was to develop, validate and compare flow injection analysis (FIA) and ultra-high-performance liquid chromatography (LC)/tandem mass spectrometry methods for the determination of imatinib in plasma from patients with chronic myeloid leukemia.

METHODS

The plasma for analysis by both methods was deproteinated by methanol containing d8-imatinib. The separation was achieved on a 1.7 μm C18 column with a linear gradient (4 mM ammonium formiate and acetonitrile, pH 3.2). FIA was performed at flow rate of 0.03 mL/min (0.1% formic acid in methanol). Multiple reaction monitoring mode on the tandem mass spectrometer (API 4000, AB Sciex) in positive ESI were used for detection.

RESULTS

The total analysis times were 3.2 (LC) and 0.75 min (FIA). Both methods were successfully validated and applied to the plasma patients samples. The limits of quantification were 4.1 and 30.8 ng/mL; imprecisions were less than 5.7% and recovery ranged between 93 and 105%, for the LC and FIA, respectively. The methods revealed an agreement with a mean difference of 1.46 ng/mL (SD 28.95 ng/mL).

CONCLUSIONS

The high-throughput methods that were developed are suitable for the therapeutic drug monitoring of imatinib in plasma. They can be used in routine clinical practice.

Для цитирования: Mičová Kateřina, Friedecký David, Faber Edgar, Polýnková Adriana, Adam Tomáš. Flow injection analysis vs. ultra high performance liquid chromatography coupled with tandem mass spectrometry for determination of imatinib in human plasma.. Clinica chimica acta; international journal of clinical chemistry. 2010. DOI: 10.1016/j.cca.2010.08.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.