A minimally invasive jet injector for intravitreal and subconjunctival injection.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Peyman Gholam A, Hosseini Kamran, Cormier Michel
Аннотация
BACKGROUND AND OBJECTIVE: To evaluate a minimally invasive injector for intravitreal and subconjunctival administration of medications.
MATERIALS AND METHODS: The device has a microneedle that communicates with an internal formulation chamber. A piercing depth-limiting flange restricts microneedle penetration to a depth of less than 1 mm and defines the location of the pars plana from the limbus. The jet injector creates a force of up to 1,000 psi, forcing the medication through the remaining sclera/choroid thickness. The device was tested in 28 enucleated rabbit eyes and 9 experimental and 4 control rabbit eyes to define jet pressure for subconjunctival and intravitreal injection.
RESULTS: Injection pressures of 76 to 156 psi were needed for subconjunctival injection and 974 psi for intravitreal injection. Clinical and histologic examinations did not reveal damage to intraocular structures.
CONCLUSION: The semi-automated jet injector facilitated intravitreal/subconjunctival injection. The microneedle-assisted jet injector minimized the risk of wet injection by anchoring the microneedle in the sclera.
MATERIALS AND METHODS: The device has a microneedle that communicates with an internal formulation chamber. A piercing depth-limiting flange restricts microneedle penetration to a depth of less than 1 mm and defines the location of the pars plana from the limbus. The jet injector creates a force of up to 1,000 psi, forcing the medication through the remaining sclera/choroid thickness. The device was tested in 28 enucleated rabbit eyes and 9 experimental and 4 control rabbit eyes to define jet pressure for subconjunctival and intravitreal injection.
RESULTS: Injection pressures of 76 to 156 psi were needed for subconjunctival injection and 974 psi for intravitreal injection. Clinical and histologic examinations did not reveal damage to intraocular structures.
CONCLUSION: The semi-automated jet injector facilitated intravitreal/subconjunctival injection. The microneedle-assisted jet injector minimized the risk of wet injection by anchoring the microneedle in the sclera.