MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3928/15428877-20111129-05

Медицинская библиотека · Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye · 2012

A minimally invasive jet injector for intravitreal and subconjunctival injection.

Peyman Gholam A, Hosseini Kamran, Cormier Michel

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

To evaluate a minimally invasive injector for intravitreal and subconjunctival administration of medications.

MATERIALS AND

METHODS

The device has a microneedle that communicates with an internal formulation chamber. A piercing depth-limiting flange restricts microneedle penetration to a depth of less than 1 mm and defines the location of the pars plana from the limbus. The jet injector creates a force of up to 1,000 psi, forcing the medication through the remaining sclera/choroid thickness. The device was tested in 28 enucleated rabbit eyes and 9 experimental and 4 control rabbit eyes to define jet pressure for subconjunctival and intravitreal injection.

RESULTS

Injection pressures of 76 to 156 psi were needed for subconjunctival injection and 974 psi for intravitreal injection. Clinical and histologic examinations did not reveal damage to intraocular structures.

CONCLUSION

The semi-automated jet injector facilitated intravitreal/subconjunctival injection. The microneedle-assisted jet injector minimized the risk of wet injection by anchoring the microneedle in the sclera.

Для цитирования: Peyman Gholam A, Hosseini Kamran, Cormier Michel. A minimally invasive jet injector for intravitreal and subconjunctival injection.. Ophthalmic surgery, lasers & imaging : the official journal of the International Society for Imaging in the Eye. 2012. DOI: 10.3928/15428877-20111129-05.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.