Metal artifact reduction in patients with total hip replacements: evaluation of clinical photon counting CT using virtual monoenergetic images.

Schreck Julian, Laukamp Kai Roman, Niehoff Julius Henning, Michael Arwed Elias, Boriesosdick Jan, Wöltjen Matthias Michael, Kröger Jan Robert, Reimer Robert P, Grunz Jan-Peter, Borggrefe Jan, Lennartz Simon

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To investigate photon-counting CT (PCCT)-derived virtual monoenergetic images (VMI) for artifact reduction in patients with unilateral total hip replacements (THR).

METHODS

Forty-two patients with THR and portal-venous phase PCCT of the abdomen and pelvis were retrospectively included. For the quantitative analysis, region of interest (ROI)-based measurements of hypodense and hyperdense artifacts, as well as of artifact-impaired bone and the urinary bladder, were conducted, and corrected attenuation and image noise were calculated as the difference of attenuation and noise between artifact-impaired and normal tissue. Two radiologists qualitatively evaluated artifact extent, bone assessment, organ assessment, and iliac vessel assessment using 5-point Likert scales.

RESULTS

VMI110keV yielded a significant reduction of hypo- and hyperdense artifacts compared to conventional polyenergetic images (CI) and the corrected attenuation closest to 0, indicating best possible artifact reduction (hypodense artifacts: CI: 237.8 ± 71.4 HU, VMI110keV: 8.5 ± 122.5 HU; p < 0.05; hyperdense artifacts: CI: 240.6 ± 40.8 HU vs. VMI110keV: 13.0 ± 110.4 HU; p < 0.05). VMI110keV concordantly provided best artifact reduction in the bone and bladder as well as the lowest corrected image noise. In the qualitative assessment, VMI110keV received the best ratings for artifact extent (CI: 2 (1-3), VMI110keV: 3 (2-4); p < 0.05) and bone assessment (CI: 3 (1-4), VMI110keV: 4 (2-5); p < 0.05), whereas organ and iliac vessel assessments were rated highest in CI and VMI70keV.

CONCLUSIONS

PCCT-derived VMI effectively reduce artifacts from THR and thereby improve assessability of circumjacent bone tissue. VMI110keV yielded optimal artifact reduction without overcorrection, yet organ and vessel assessments at that energy level and higher were impaired by loss of contrast.

CLINICAL RELEVANCE STATEMENT: PCCT-enabled artifact reduction is a feasible method for improving assessability of the pelvis in patients with total hip replacements at clinical routine imaging.

KEY POINTS: • Photon-counting CT-derived virtual monoenergetic images at 110 keV yielded best reduction of hyper- and hypodense artifacts, whereas higher energy levels resulted in artifact overcorrection. • The qualitative artifact extent was reduced best in virtual monoenergetic images at 110 keV, facilitating an improved assessment of the circumjacent bone. • Despite significant artifact reduction, assessment of pelvic organs as well as vessels did not profit from energy levels higher than 70 keV, due to the decline in image contrast.

Для цитирования: Schreck Julian, Laukamp Kai Roman, Niehoff Julius Henning, Michael Arwed Elias, Boriesosdick Jan, Wöltjen Matthias Michael, Kröger Jan Robert, Reimer Robert P, Grunz Jan-Peter, Borggrefe Jan, Lennartz Simon. Metal artifact reduction in patients with total hip replacements: evaluation of clinical photon counting CT using virtual monoenergetic images.. European radiology. 2023. DOI: 10.1007/s00330-023-09879-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.