DOI 10.1016/j.juro.2015.02.083

Медицинская библиотека · The Journal of urology · 2015

Preventing Urethral Trauma from Inadvertent Inflation of Catheter Balloon in the Urethra during Catheterization: Evaluation of a Novel Safety Syringe after Correlating Trauma with Urethral Distension and Catheter Balloon Pressure.

Davis Niall F, Mooney Rory O'C, Cunnane Conor V, Cunnane Eoghan M, Thornhill John A, Walsh Michael T

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: We investigated urethral diametric strain and threshold maximum inflation pressure for rupture during inadvertent inflation of a catheter anchoring balloon in the urethra. In addition, we evaluated a novel safety device to prevent trauma based on these parameters.

MATERIALS AND

METHODS

Inflation of a urethral catheter anchoring balloon was performed in the bulbar urethra of 21 ex vivo porcine models using 16Fr catheters. Urethral trauma was assessed with retrograde urethrography. Urethral rupture was correlated with internal urethral diametric strain and maximal urethral pressure threshold values in kPa. Urethral catheters were then inflated in the bulbar urethras of 7 fresh male cadavers using a standard syringe and a prototype syringe. The plunger of the standard syringe was depressed until opposing resistance pressure generated by the urethra prevented further inflation of the anchoring balloon. The plunger of the prototype safety syringe was depressed until sterile water in the syringe decanted through an activated safety threshold pressure valve.

RESULTS

Retrograde urethrography demonstrated that porcine urethral rupture consistently occurred at an internal urethral diametric strain greater than 40% and a maximum inflation pressure greater than 150 kPa. The mean ± SD maximum human urethral threshold inflation pressure required to activate the safety prototype syringe pressure valve was 153 ± 3 kPa. In comparison, maximum inflation pressure was significantly greater using the standard syringe than the activated prototype syringe (mean 452 ± 188 kPa, p <0.001).

CONCLUSIONS

Internal urethral diametric strain and threshold maximum inflation pressures are important parameters for designing a safer urethral catheter system with lower intrinsic threshold inflation pressures.

Для цитирования: Davis Niall F, Mooney Rory O'C, Cunnane Conor V, Cunnane Eoghan M, Thornhill John A, Walsh Michael T. Preventing Urethral Trauma from Inadvertent Inflation of Catheter Balloon in the Urethra during Catheterization: Evaluation of a Novel Safety Syringe after Correlating Trauma with Urethral Distension and Catheter Balloon Pressure.. The Journal of urology. 2015. DOI: 10.1016/j.juro.2015.02.083.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.