Quantification of histochemical stains using whole slide imaging: development of a method and demonstration of its usefulness in laboratory quality control.
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Gray Allan, Wright Alex, Jackson Pete, Hale Mike, Treanor Darren
Аннотация
AIMS: Histochemical staining of tissue is a fundamental technique in tissue diagnosis and research, but it suffers from significant variability. Efforts to address this include laboratory quality controls and quality assurance schemes, but these rely on subjective interpretation of stain quality, are laborious and have low reproducibility. We aimed (1) to develop a method for histochemical stain quantification using whole slide imaging and image analysis and (2) to demonstrate its usefulness in measuring staining variation.
METHODS: A method to quantify the individual stain components of histochemical stains on virtual slides was developed. It was evaluated for repeatability and reproducibility, then applied to control sections of an appendix to quantify H&E staining (H/E intensities and H:E ratio) between automated staining machines and to measure differences between six regional diagnostic laboratories.
RESULTS: The method was validated with <0.5% variation in H:E ratio measurement when using the same scanner for a batch of slides (ie, it was repeatable) but was not highly reproducible between scanners or over time, where variation of 7% was found. Application of the method showed H:E ratios between three staining machines varied from 0.69 to 0.93, H:E ratio variation over time was observed. Interlaboratory comparison demonstrated differences in H:E ratio between regional laboratories from 0.57 to 0.89.
CONCLUSIONS: A simple method using whole slide imaging can be used to quantify and compare histochemical staining. This method could be deployed in routine quality assurance and quality control. Work is needed on whole slide imaging devices to improve reproducibility.
METHODS: A method to quantify the individual stain components of histochemical stains on virtual slides was developed. It was evaluated for repeatability and reproducibility, then applied to control sections of an appendix to quantify H&E staining (H/E intensities and H:E ratio) between automated staining machines and to measure differences between six regional diagnostic laboratories.
RESULTS: The method was validated with <0.5% variation in H:E ratio measurement when using the same scanner for a batch of slides (ie, it was repeatable) but was not highly reproducible between scanners or over time, where variation of 7% was found. Application of the method showed H:E ratios between three staining machines varied from 0.69 to 0.93, H:E ratio variation over time was observed. Interlaboratory comparison demonstrated differences in H:E ratio between regional laboratories from 0.57 to 0.89.
CONCLUSIONS: A simple method using whole slide imaging can be used to quantify and compare histochemical staining. This method could be deployed in routine quality assurance and quality control. Work is needed on whole slide imaging devices to improve reproducibility.
Предметные рубрики MeSH
AppendixAutomation, LaboratoryClinical Laboratory TechniquesColoring AgentsEosine Yellowish-(YS)HematoxylinHumansImage Interpretation, Computer-AssistedLaboratoriesLaboratory Proficiency TestingObserver VariationPredictive Value of TestsQuality ControlQuality Indicators, Health CareReproducibility of ResultsSpecimen HandlingStaining and Labeling