Larson Erik R, Porterfield John E, Sagar Sandeep, Marmol-Velez Juan, Panday Manoj, Escobedo Daniel, Michalek Joel, Ouyang Yongjian, Valvano Jonathan W, Pearce John A, Feldman Marc D
Аннотация
BACKGROUND: Implantable cardioverter-defibrillators monitor intracardiac electrograms (EGMs) to discriminate between ventricular and supraventricular tachycardias. The incidence of inappropriate shocks remains high because of misclassification of the tachycardia in an otherwise hemodynamically stable individual. Coupling EGMs with an assessment of left ventricular (LV) stroke volume (SV) could help in gauging hemodynamics during an arrhythmia and reducing inappropriate shocks.
OBJECTIVE: The purpose of this study was to use the admittance method to accurately derive LV SV.
METHODS: Ultrasonic flow probe and LV endocardial crystals were used in canines (n = 12) as the standard for LV SV. Biventricular pacing leads were inserted to obtain admittance measurements. A tetrapolar, complex impedance measurement was made between the Bi-V leads. The real and imaginary components of impedance were used to discard the myocardial component from the blood component to derive instantaneous blood conductance (Gb). Alterations in SV were measured during right ventricular pacing, dopamine infusion, and inferior vena cava occlusion.
RESULTS: Gb tracks steady-state changes in SV more accurately than traditional magnitude (ie, |Y|, without removal of the muscle signal) during right ventricular pacing and dopamine infusion (P = .004). Instantaneous LV volume also was tracked more accurately by Gb than ∣Y∣ in the subset of subjects that underwent inferior vena cava occlusions (n = 5, P = .025). Finite element modeling demonstrates that admittance shifts more sensitivity of the measurement to the LV blood chamber as the mechanism for improvement (see Online Appendix).
CONCLUSION: Monitoring LV SV is possible using the admittance method with biventricular pacing leads. The technique could be piggybacked to complement EGMs to determine if arrhythmias are hemodynamically unstable.
OBJECTIVE: The purpose of this study was to use the admittance method to accurately derive LV SV.
METHODS: Ultrasonic flow probe and LV endocardial crystals were used in canines (n = 12) as the standard for LV SV. Biventricular pacing leads were inserted to obtain admittance measurements. A tetrapolar, complex impedance measurement was made between the Bi-V leads. The real and imaginary components of impedance were used to discard the myocardial component from the blood component to derive instantaneous blood conductance (Gb). Alterations in SV were measured during right ventricular pacing, dopamine infusion, and inferior vena cava occlusion.
RESULTS: Gb tracks steady-state changes in SV more accurately than traditional magnitude (ie, |Y|, without removal of the muscle signal) during right ventricular pacing and dopamine infusion (P = .004). Instantaneous LV volume also was tracked more accurately by Gb than ∣Y∣ in the subset of subjects that underwent inferior vena cava occlusions (n = 5, P = .025). Finite element modeling demonstrates that admittance shifts more sensitivity of the measurement to the LV blood chamber as the mechanism for improvement (see Online Appendix).
CONCLUSION: Monitoring LV SV is possible using the admittance method with biventricular pacing leads. The technique could be piggybacked to complement EGMs to determine if arrhythmias are hemodynamically unstable.