Imaging Optimization of Temporal Bones With Cochlear Implant Using a High-resolution Cone Beam CT and the Corresponding Effective Dose.

Zou Jing, Koivisto Juha, Lähelmä Jaakko, Aarnisalo Antti, Wolff Jan, Pyykkö Ilmari

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the impact of tube voltage, tube current, pulse number, and magnification factor on the image quality of a novel experimental set-up and the corresponding radiation.

MATERIALS AND

METHODS

Six human temporal bones with cochlear implant were imaged using various tube voltages, tube currents, pulse numbers, and magnification. The effect of radiation was evaluated using a metaloxide semiconductor field-effect transistor (MOSFET) dosimeter device on an anthropomorphic RANDO RAN102 male head phantom. A copper and aluminum combination filter was used for hardware filtration.

RESULTS

Overall, 900 frames, 11 mA, and 88 kV provided the best image quality. In temporal bones imaged with the optimized parameters, the cochlea, osseous spiral lamina, modiolus, stapes, round window niche, and oval window landmarks were demonstrated with anatomic structures still fully assessable in all parts and acceptable image quality. The most dominant contributor to the effective dose was bone marrow (36%-37 %) followed by brain (34%-36%), remainder tissues (12%), extra-thoracic airways (7%), and oral mucosa (5%).

CONCLUSIONS

By increasing the number of frames, the image quality of the inner ear details obtained using the novel cone-beam computed tomography improved.

Для цитирования: Zou Jing, Koivisto Juha, Lähelmä Jaakko, Aarnisalo Antti, Wolff Jan, Pyykkö Ilmari. Imaging Optimization of Temporal Bones With Cochlear Implant Using a High-resolution Cone Beam CT and the Corresponding Effective Dose.. The Annals of otology, rhinology, and laryngology. 2014. DOI: 10.1177/0003489414565004.