MedElevator Войти в исследования
← Поиск публикаций

DOI 10.7874/kja.2014.18.2.93

Медицинская библиотека · Korean journal of audiology · 2014

A case of the vibrant soundbridge stapes coupler in patients with mixed hearing loss.

Park Ah Young, Jeon Ju Hyun, Moon In Seok, Choi Jae Young

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

The Vibrant Soundbridge (VSB) with stapes clip coupler placement at the stapes head has been used successfully to treat mixed hearing loss. Coupling between the floating mass transducer of the VSB and the stapes head is technically less demanding than incus vibroplasty and is more likely to generate a positive outcome without significantly changing residual hearing or resulting in medical or surgical complications. A 65-year-old man with bilateral mixed hearing loss and chronic otitis media underwent vibroplasty with a stapes clip coupler. Speech discrimination scores in both quiet and noise environments showed better functional gain with the VSB than with the use of a conventional hearing aid. The results of the present case show the feasibility of implanting a VSB with a stapes coupler in patients with mixed hearing loss due to chronic otitis media.

Для цитирования: Park Ah Young, Jeon Ju Hyun, Moon In Seok, Choi Jae Young. A case of the vibrant soundbridge stapes coupler in patients with mixed hearing loss.. Korean journal of audiology. 2014. DOI: 10.7874/kja.2014.18.2.93.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.