MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000000577

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2014

Comparison of two different titanium couplers for an active middle ear implant.

Schwab Burkard, Salcher Rolf, Teschner Magnus

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare surgical methods, functional gain, and speech discrimination using two different coupling methods for an active middle ear implant. Of several couplers enabling placement of the active element at various locations, two function directly at a cochlear membrane, bypassing a missing or malformed ossicular chain. This study evaluates whether either of these methods is more beneficial.

STUDY

DESIGN

Retrospective case review.

SETTING

ENT surgical clinic.

PATIENTS: Forty-seven German-speaking patients with moderate to severe mixed hearing loss.

INTERVENTIONS: Records of patients implanted with either a round window (RW) or oval window (OW) coupler and active implant were examined. Preoperative and postoperative bone and air-conduction thresholds, auditory gain, and speech perception were compared.

MAIN OUTCOME MEASURES: Functional gain, Freiburger monosyllables in quiet.

RESULTS

The range of hearing benefit shown by functional gain in patients implanted with the RW coupler (median) was between 22.5 dB (at 0.25 kHz) and 52.5 dB (2 and 3 kHz). In the OW group, improvement was similar, ranging from 21 dB (at 8 kHz) to 50 dB (1 and 2 kHz). Patients in both groups showed a similar improvement in speech recognition. Median preoperative unaided word recognition was 0% at 60 dB HL for both patient groups, improved postoperatively in both groups to median 85% correct at 65 dB HL and 95% at 80 dB HL.

CONCLUSION

Placement of an active middle ear implant using the RW and the OW coupler was found to be safe, although the surgical methods differ. Safety and efficacy of both couplers present no significant differences.

Для цитирования: Schwab Burkard, Salcher Rolf, Teschner Magnus. Comparison of two different titanium couplers for an active middle ear implant.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2014. DOI: 10.1097/mao.0000000000000577.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.