Accuracy and precision of in vitro EOS imaging compared to digital radiographs in the measurement of intramedullary lengthening.

Al Ramlawi Akram, Nahm Nickolas, Abdou Michael, Salem Hytham S, Herzenberg John E, McClure Philip

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Accurate limb length measurement is essential during limb lengthening procedures to prevent neurologic and musculoskeletal complications. Magnetic intramedullary lengthening nails (MILN) rely on radiographs for measurement, but the optimal radiographic technique is not established. This study evaluated the precision and accuracy of EOS imaging versus digital radiography, including calibration techniques and measurement techniques for assessing intramedullary lengthening.

METHODS

An MILN was inserted into a Sawbones femur, and lengthening measurements from digital X-ray and EOS imaging were compared to external remote control (ERC) results. Measurements were taken with and without calibration using a magnification ball, nail width, or female nail length. Four observers measured the distraction gap, spindle length, full nail length, and male nail length at various hip flexion angles. Precision and accuracy were calculated for each technique.

RESULTS

From 576 measurements, EOS imaging demonstrated significantly higher accuracy than digital radiography (83.6% vs. 73.5% absolute accuracy; p < 0.001) when non-calibrated values were included. EOS maintained superior accuracy at 30 mm, 50 mm, and 70 mm of lengthening (p < 0.05), with no difference at 10 mm. Calibration did not affect EOS accuracy, whereas digital radiography was less accurate without calibration (p < 0.001). Distraction gap measurement was the most accurate for both modalities, and 0° hip flexion provided the highest precision. Overall, EOS was more precise (79.4% vs. 71.7% precision rates; p < 0.001).

CONCLUSION

EOS imaging is recommended for limb lengthening measurements due to its superior precision and accuracy. For digital radiographs, careful calibration is essential to achieve accurate measurements.

Для цитирования: Al Ramlawi Akram, Nahm Nickolas, Abdou Michael, Salem Hytham S, Herzenberg John E, McClure Philip. Accuracy and precision of in vitro EOS imaging compared to digital radiographs in the measurement of intramedullary lengthening.. Skeletal radiology. 2025. DOI: 10.1007/s00256-025-04879-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.