Magnetic resonance in medicine · 2022
Phase-cycled balanced SSFP imaging for non-contrast-enhanced functional lung imaging.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ilicak Efe, Ozdemir Safa, Schad Lothar R, Weis Meike, Schoenberg Stefan O, Zöllner Frank G, Zapp Jascha
Аннотация
PURPOSE: To introduce phase-cycled balanced SSFP (bSSFP) acquisition as an alternative in Fourier decomposition MRI for improved robustness against field inhomogeneities.
METHODS: Series 2D dynamic lung images were acquired in 5 healthy volunteers at 1.5 T and 3 T using bSSFP sequence with multiple RF phase increments and compared with conventional single RF phase increment acquisitions. The approach was evaluated based on functional map homogeneity analysis, while ensuring image and functional map quality by means of SNR and contrast-to-noise ratio analyses.
RESULTS: At both field strengths, functional maps obtained with phase-cycled acquisitions displayed improved robustness against local signal losses compared with single-phase acquisitions. The coefficient of variation (mean ± SD, across volunteers) measured in the ventilation maps resulted in 29.7 ± 2.6 at 1.5 T and 37.5 ± 3.1 at 3 T for phase-cycled acquisitions, compared with 39.9 ± 5.2 at 1.5 T and 49.5 ± 3.7 at 3 T for single-phase acquisitions, indicating a significant improvement ( p<0.05$$ p<0.05 $$ ) in ventilation map homogeneity.
CONCLUSIONS: Phase-cycled bSSFP acquisitions improve robustness against field inhomogeneity artifacts and significantly improve ventilation map homogeneity at both field strengths. As such, phase-cycled bSSFP may serve as a robust alternative in lung function assessments.
METHODS: Series 2D dynamic lung images were acquired in 5 healthy volunteers at 1.5 T and 3 T using bSSFP sequence with multiple RF phase increments and compared with conventional single RF phase increment acquisitions. The approach was evaluated based on functional map homogeneity analysis, while ensuring image and functional map quality by means of SNR and contrast-to-noise ratio analyses.
RESULTS: At both field strengths, functional maps obtained with phase-cycled acquisitions displayed improved robustness against local signal losses compared with single-phase acquisitions. The coefficient of variation (mean ± SD, across volunteers) measured in the ventilation maps resulted in 29.7 ± 2.6 at 1.5 T and 37.5 ± 3.1 at 3 T for phase-cycled acquisitions, compared with 39.9 ± 5.2 at 1.5 T and 49.5 ± 3.7 at 3 T for single-phase acquisitions, indicating a significant improvement ( p<0.05$$ p<0.05 $$ ) in ventilation map homogeneity.
CONCLUSIONS: Phase-cycled bSSFP acquisitions improve robustness against field inhomogeneity artifacts and significantly improve ventilation map homogeneity at both field strengths. As such, phase-cycled bSSFP may serve as a robust alternative in lung function assessments.