European journal of radiology · 2022
Inaccurate table height setting affects the organ-specific radiation dose in computed tomography.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ishita Yuya, Urikura Atsushi, Yoshida Tsukasa, Takiguchi Keisuke, Ikegaya Masami
Аннотация
PURPOSE: We aimed to prove that the locally absorbed doses in tissues and organs are affected by inaccurate table height in computed tomography.
MATERIALS AND METHODS: We compared the volume CT dose index (CTDIvol) and the absorbed doses using an anthropomorphic phantom combined with a breast phantom. The phantom was set at the gantry center, from which the table height was changed every 20 mm between-40 mm and 40 mm. Data acquisition was performed using auto table height correction (AHC) for each table height. The CTDIvol was obtained from the CT console and the tube current value for each image slice (DICOM tag: 0018, 1151). The absorbed dose was measured by a glass dosimeter that was implanted at various positions in the phantom.
RESULTS: The tube current values in the lung were lower at a table height of + 40 mm than those at other heights. The CTDIvol was slightly lower at + 40 mm than at the center (12.78 mGy vs. 13.42 mGy, p < 0.05). The CTDIvol values were almost the same at the other table heights (13.30-13.40 mGy). The absorbed doses at the lens and mammary gland were significantly different from those at the gantry center (-27.27%-17.77% and -24.31%-12.83%, respectively). Compared with the center, both the lens and mammary gland had higher absorbed doses at a table height of -40 mm.
CONCLUSION: The absorbed dose was affected by the table height, but the CTDIvol was maintained by AHC. The operator should appropriately position patients even when using AHC.
MATERIALS AND METHODS: We compared the volume CT dose index (CTDIvol) and the absorbed doses using an anthropomorphic phantom combined with a breast phantom. The phantom was set at the gantry center, from which the table height was changed every 20 mm between-40 mm and 40 mm. Data acquisition was performed using auto table height correction (AHC) for each table height. The CTDIvol was obtained from the CT console and the tube current value for each image slice (DICOM tag: 0018, 1151). The absorbed dose was measured by a glass dosimeter that was implanted at various positions in the phantom.
RESULTS: The tube current values in the lung were lower at a table height of + 40 mm than those at other heights. The CTDIvol was slightly lower at + 40 mm than at the center (12.78 mGy vs. 13.42 mGy, p < 0.05). The CTDIvol values were almost the same at the other table heights (13.30-13.40 mGy). The absorbed doses at the lens and mammary gland were significantly different from those at the gantry center (-27.27%-17.77% and -24.31%-12.83%, respectively). Compared with the center, both the lens and mammary gland had higher absorbed doses at a table height of -40 mm.