PURPOSE: This study aimed to evaluate the intrafractional prostate motion captured during gated magnetic resonance imaging (MRI)-guided online adaptive radiotherapy for prostate cancer and analyze its impact on the delivered dose as well as the effect of gating.

METHODS

Sagittal 2D cine-MRI scans were acquired at 4 Hz during treatment at a ViewRay MRIdian (ViewRay Inc., Oakwood Village, OH, USA) MR linac. Prostate shifts in anterior-posterior (AP) and superior-inferior (SI) directions were extracted separately. Using the static dose cloud approximation, the planned fractional dose was shifted according to the 2D gated motion (residual motion in gating window) to estimate the delivered dose by superimposing and averaging the shifted dose volumes. The dose of a hypothetical non-gated delivery was reconstructed similarly using the non-gated motion. For the clinical target volume (CTV), rectum, and bladder, dose-volume histogram parameters of the planned and reconstructed doses were compared.

RESULTS

In total, 174 fractions (15.7 h of cine-MRI) from 10 patients were evaluated. The average (±1 σ) non-gated prostate motion was 0.6 ± 1.0 mm in the AP and 0.0 ± 0.6 mm in the SI direction with respect to the centroid position of the gating boundary. 95% of the shifts were within [-3.5, 2.7] mm in the AP and [-2.9, 3.2] mm in the SI direction. For the gated treatment and averaged over all fractions, CTV D98% decreased by less than 2% for all patients. The rectum and the bladder D2% increased by less than 3% and 0.5%, respectively. Doses reconstructed for gated and non-gated delivery were similar for most fractions.

CONCLUSION

A pipeline for extraction of prostate motion during gated MRI-guided radiotherapy based on 2D cine-MRI was implemented. The 2D motion data enabled an approximate estimation of the delivered dose. For the majority of fractions, the benefit of gating was negligible, and clinical dosimetric constraints were met, indicating safety of the currently adopted gated MRI-guided treatment workflow.

Для цитирования: Xiong Yuqing, Rabe Moritz, Nierer Lukas, Kawula Maria, Corradini Stefanie, Belka Claus, Riboldi Marco, Landry Guillaume, Kurz Christopher. Assessment of intrafractional prostate motion and its dosimetric impact in MRI-guided online adaptive radiotherapy with gating.. Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]. 2022. DOI: 10.1007/s00066-022-02005-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.