Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2012

Diagnostic performance of two highly multiplexed respiratory virus assays in a pediatric cohort.

Forman Michael S, Advani Sonali, Newman Christina, Gaydos Charlotte A, Milstone Aaron M, Valsamakis Alexandra

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Rapid detection of respiratory viruses is important for management and infection control in hospitalized patients. Multiplex nucleic acid tests (NATs) have begun to replace conventional methods as gold standards for respiratory virus detection.

OBJECTIVE

To compare the performance of two large multiplex NATS, ResPlex II (RPII) and Respiratory Virus Surveillance kit with electrospray ionization mass spectrometry (RVS/MS) using nasopharyngeal aspirates (NPAs) from hospitalized children who had been tested previously with conventional methods.

STUDY

DESIGN

Stored residual NPAs (N=306) were tested concomitantly by RPII and RVS/MS. Alternate NATs were used to adjudicate discordant results.

RESULTS

More viruses were detected with multiplex NATs (RPII, 110; RVS/MS, 109) than conventional assays (86); diagnostic gain was primarily for fastidious viruses (coronaviruses and enteroviruses [EVs]/human rhinoviruses [HRVs]). Total positive and negative agreement between the multiplex NATs for all viruses detected was quite high (86% positive agreement, 99% negative agreement). Most individual viruses were detected with fairly equivalent accuracy by the multiplex NATs, except for adenoviruses (RPII sensitivity 40%) and human metapneumovirus (RVS/MS sensitivity 42%). RPII had the advantage of detecting EVs and HRVs, however, it demonstrated considerable EV/HRV cross-reactivity (29 HRV-positive specimens by real-time PCR were positive for EV by RPII and 21 specimens positive for HRV only by RT-PCR were dual positive for EV/HRV by RPII). RPII also had reduced sensitivity for HRV detection (in 36 specimens, HRV was detected by RT-PCR but not by RPII).

CONCLUSIONS

Both multiplex NATs were promising, but had notable limitations.

Для цитирования: Forman Michael S, Advani Sonali, Newman Christina, Gaydos Charlotte A, Milstone Aaron M, Valsamakis Alexandra. Diagnostic performance of two highly multiplexed respiratory virus assays in a pediatric cohort.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2012. DOI: 10.1016/j.jcv.2012.06.019.