Design study of a miniaturized displacement transducer (MDT) for an active middle ear implant system.
Аннотация
People suffering from moderate to severe hearing loss can be treated with active middle ear implants. A new approach in this field is to implant an electromechanical transducer onto the round window membrane in order to improve coupling and be able to treat patients with middle-ear problems. In this paper the design study for a miniaturized displacement transducer (MDT) for the round window is presented. Based on a requirement analysis, the basic principle and analytical modeling of the actuator is shown. A parameter variation study results in an optimized actuator configuration that is able to generate an amplification of 110 dB SPL theoretically. As a next step this actuator has to be manufactured and tested.