In vivo (Athens, Greece) · 2023

Relative Biological Effectiveness Values of Spot-scanning Proton Beam Therapy at Shonan Kamakura General Hospital.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Shiba Shintaro, Yamanaka Masashi, Matsumoto Kazuki, Yamano Akihiro, Shimo Takahiro, Suzuki Shunsuke, Yagihashi Takayuki, Nitta Kazunori, Ohno Tatsuya, Tokuuye Koichi, Omura Motoko

Аннотация

BACKGROUND/AIM: This study aimed to confirm the relative biological effectiveness (RBE) values of the proton beam therapy (PBT) system installed in Shonan Kamakura General Hospital.

MATERIALS AND METHODS: Clonogenic cell-survival assays were performed with a human salivary gland (HSG) cell line, a human tongue squamous-cell carcinoma cell line (SAS), and a human osteosarcoma cell line (MG-63). Cells were irradiated with proton beams and X-rays with different doses (1.8, 3.6, 5.5, and 7.3 Gy for proton beams, and 2, 4, 6, and 8 Gy for X-rays). Proton beam irradiation used spot-scanning methods and three different depths (at the proximal, center, and distal sides of the spread-out Bragg peak). RBE values were obtained from a comparison of the dose that resulted in a surviving fraction of 10% (D10).

RESULTS: D10 of proton beams at the proximal, center, and distal sides and X-rays in HSG were 4.71, 4.71, 4.51, and 5.25 Gy, respectively; those in SAS were 5.08, 5.04, 5.01, and 5.59 Gy, respectively; and those in MG-63 were 5.36, 5.42, 5.12, and 6.06 Gy, respectively. The RBE10 values at the proximal, center, and distal sides in HSG were 1.11, 1.11, and 1.16 respectively; those in SAS were 1.10, 1.11, and 1.12, respectively; and those in MG-63 were 1.13, 1.12, and 1.18, respectively.

CONCLUSION: RBE10 values of 1.10-1.18 were confirmed by in vitro experiments using the PBT system. These results are considered acceptable for clinical use in terms of therapeutic efficacy and safety.