Accuracy of non-invasive smartphone bilirubin measurements on neonates before and during phototherapy.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Zdraveska Nikolina, Kostovski Aco, Reich Thilo, Hill Deborah K, Gierman Lobke M, Bahr Timothy M

Аннотация

UNLABELLED: Previous publications have reported the accuracy of the Picterus Jaundice Pro (Picterus JP), a smartphone-based medical device app for screening neonatal jaundice prior to phototherapy. The purpose of this study was to evaluate the accuracy of the Picterus JP app before and during phototherapy. Between April 2025 and April 2026, newborns whose total serum bilirubin concentration (TSB) reached or exceeded the American Academy of Pediatrics phototherapy thresholds were enrolled in the study after written informed consent was obtained from their parent or legal guardian. A Picterus JP scan was obtained from the chest of the newborn (within 60 min of the TSB blood sample collection), a light-occlusive patch was placed on the newborn's chest, and phototherapy was initiated. Then, up to five Picterus JP scans per newborn were obtained from the newborn's chest, from the region covered by the phototherapy patch, within one hour of the blood collection for subsequent TSBs. We calculated diagnostic accuracy statistics to summarize the accuracy of Picterus JP. A total of 137 paired Picterus JP-TSB measurements were analyzed. Overall, Picterus JP correlated well with TSB (r = 0.720; bias - 38.5 µmol/L; RMSE 73.7 µmol/L). Restricted to a phototherapy course of ≤ 3 days, correlation was strong before phototherapy (n = 45 measurements; r = 0.849; bias - 46.5 µmol/L; RMSE 65.3 µmol/L) and weaker during phototherapy (n = 83 measurements; r = 0.520; bias - 40.7 µmol/L; RMSE 79.6 µmol/L), although the magnitude of bias and RMSE was similar in both periods. The relative (fractional) decline in Picterus JP during phototherapy correlated well with the relative decline in TSB (n = 97 paired change measurements; r = 0.752), and the direction of the bilirubin trend agreed between methods in 83% of assessed trajectories. A fixed offset correction reduced the bias during phototherapy from - 40.7 to + 5.8 µmol/L.

CONCLUSION: Although the correlation between Picterus JP results and TSB measurements decreased during phototherapy, the direction of change in Picterus JP results almost always agreed with the direction of change in associated TSB measurements. This suggests that Picterus JP may be useful for monitoring bilirubin trends and treatment response during phototherapy, while also providing non-invasive bilirubin assessment before treatment initiation.

WHAT IS KNOWN: • Total serum bilirubin (TSB) is the reference standard for guiding phototherapy but requires invasive blood draws; smartphone-based tools are validated mainly before phototherapy.

WHAT IS NEW: • Picterus JP tracked bilirubin trends during phototherapy despite reduced correlation; an offset correction reduced bias, extending validation to severe hyperbilirubinemia.